Improved volumetric measurement of brain structure with a distortion correction procedure using an ADNI phantom
Authors:
- Norihide MaikusaNMNorihide Maikusa
- Fumio YamashitaFYFumio Yamashita
- Kenichiro TanakaKTKenichiro Tanaka
- Osamu AbeOAOsamu Abe
- Atsushi KawaguchiAKAtsushi Kawaguchi
- Hiroyuki KabasawaHKHiroyuki Kabasawa
- Shoma ChibaSCShoma Chiba
- Akihiro KasaharaAKAkihiro Kasahara
- Nobuhisa KobayashiNKNobuhisa Kobayashi
- Tetsuya YuasaTYTetsuya Yuasa
- Noriko SatoNSNoriko Sato
- Hiroshi MatsudaHMHiroshi Matsuda
- 28 May 2013
- journal article
- Published by Wiley in Medical Physics
- Vol. 40 (6Part1) , 062303
Abstract
Purpose: Serial magnetic resonance imaging (MRI) images acquired from multisite and multivendor MRI scanners are widely used in measuring longitudinal structural changes in the brain. Precise and accurate measurements are important in understanding the natural progression of neurodegenerative disorders such as Alzheimer's disease. However, geometric distortions in MRI images decrease the accuracy and precision of volumetric or morphometric measurements. To solve this problem, the authors suggest a commercially available phantom‐based distortion correction method that accommodates the variation in geometric distortion within MRI images obtained with multivendor MRI scanners.Methods: The authors' method is based on image warping using a polynomial function. The method detects fiducial points within a phantom image using phantom analysis software developed by the Mayo Clinic and calculates warping functions for distortion correction. To quantify the effectiveness of the authors' method, the authors corrected phantom images obtained from multivendor MRI scanners and calculated the root‐mean‐square (RMS) of fiducial errors and the circularity ratio as evaluation values. The authors also compared the performance of the authors' method with that of a distortion correction method based on a spherical harmonics description of the generic gradient design parameters. Moreover, the authors evaluated whether this correction improves the test–retest reproducibility of voxel‐based morphometry in human studies.Results: A Wilcoxon signed‐rank test with uncorrected and corrected images was performed. The root‐mean‐square errors and circularity ratios for all slices significantly improved (p < 0.0001) after the authors' distortion correction. Additionally, the authors' method was significantly better than a distortion correction method based on a description of spherical harmonics in improving the distortion of root‐mean‐square errors (p < 0.001 and 0.0337, respectively). Moreover, the authors' method reduced the RMS error arising from gradient nonlinearity more than gradwarp methods. In human studies, the coefficient of variation of voxel‐based morphometry analysis of the whole brain improved significantly from 3.46% to 2.70% after distortion correction of the whole gray matter using the authors' method (Wilcoxon signed‐rank test, p < 0.05).Conclusions: The authors proposed a phantom‐based distortion correction method to improve reproducibility in longitudinal structural brain analysis using multivendor MRI. The authors evaluated the authors' method for phantom images in terms of two geometrical values and for human images in terms of test–retest reproducibility. The results showed that distortion was corrected significantly using the authors' method. In human studies, the reproducibility of voxel‐based morphometry analysis for the whole gray matter significantly improved after distortion correction using the authors' method.Keywords
This publication has 26 references indexed in Scilit:
- Automatic classification of patients with Alzheimer's disease from structural MRI: A comparison of ten methods using the ADNI databaseNeuroImage, 2011
- 3D maps localize caudate nucleus atrophy in 400 Alzheimer’s disease, mild cognitive impairment, and healthy elderly subjectsNeurobiology of Aging, 2010
- Gradient distortions in MRI: Characterizing and correcting for their effects on SIENA-generated measures of brain volume changeNeuroImage, 2010
- Comparison of phantom and registration scaling corrections using the ADNI cohortNeuroImage, 2009
- Structural MRI biomarkers for preclinical and mild Alzheimer's diseaseHuman Brain Mapping, 2009
- The Alzheimer's disease neuroimaging initiative (ADNI): MRI methodsJournal of Magnetic Resonance Imaging, 2008
- Alzheimer's disease diagnosis in individual subjects using structural MR images: Validation studiesNeuroImage, 2008
- Automatic classification of MR scans in Alzheimer's diseaseBrain, 2008
- Information criteria for astrophysical model selectionMonthly Notices of the Royal Astronomical Society: Letters, 2007
- Estimating the Dimension of a ModelThe Annals of Statistics, 1978