Point‐of‐care glucose monitoring on the neonatal unit
Authors:
- 23 November 2011
- journal article
- Published by Wiley in Journal of Paediatrics and Child Health
- Vol. 48 (4) , 342-346
Abstract
Aim: This study aims to compare the accuracy and precision of the Nova StatStrip glucometer against the Radiometer ABL gas analyser. Based on the results, to establish if the Nova StatStrip glucometer could be adopted as a reliable alternative for near‐patient glucose monitoring on the neonatal unit.Methods: Seven hundred twenty‐eight paired samples were collected prospectively from babies on a neonatal intensive care unit. Analytical performance of the Nova StatStrip glucometer was assessed based on the ISO 15197 criteria and the American Diabetic Association standards. Its performance compared with the Radiometer ABL gas analyser was assessed statistically using Bland–Altman analysis and clinically by use of an error grid.Results: A percentage of 98.8 of StatStrip values less than 4.2 mmol/L and 97.9% of values greater than 4.2 mmol/L met the ISO criteria. Bland–Altman analysis showed good correlation between the readings. An error grid showed that most infants would be appropriately managed for hypoglycaemic episodes as per local guidelines.Conclusions: The Nova StatStrip performed well on statistical analysis compared with the Radiometer. Very few hypoglycaemic patients would be missed using the Nova StatStrip glucometer. We would recommend its use on our unit.Keywords
This publication has 10 references indexed in Scilit:
- Acidification of Blood Is Superior to Sodium Fluoride Alone as an Inhibitor of GlycolysisClinical Chemistry, 2009
- Neurologic outcome in patients with MRI pattern of damage typical for neonatal hypoglycemiaBrain and Development, 2007
- Glucose Measurement: Confounding Issues in Setting Targets for Inpatient ManagementDiabetes Care, 2007
- Evaluation of “point of care” devices in the measurement of low blood glucose in neonatal practiceArchives of Disease in Childhood - Fetal and Neonatal Edition, 2004
- Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care TestingArchives of Pathology & Laboratory Medicine, 2000
- Controversies Regarding Definition of Neonatal Hypoglycemia: Suggested Operational ThresholdsPediatrics, 2000
- Epidemiological Analysis Defining Concurrent Outbreaks of Serratia marcescens and Methicillin-Resistant Staphylococcus aureus in a Neonatal Intensive-Care UnitInfection Control & Hospital Epidemiology, 1998
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Glucose determinations in plasma and serum: potential error related to increased hematocrit.Clinical Chemistry, 1982