This study reports the intermediate-term results of four patients from a series eight patients who have had an insertion of a new complete distal radioulnar joint (DRUJ) prosthesis. A retrospective review was performed for all patients who underwent DRUJ joint replacement with the STABILITY Sigmoid Notch Total DRUJ System between the years of 2003 and 2008. To be included in this report, all patients had to have more than 24 months of follow-up and hand examination by the senior author. From the eight patients with this procedure, four patients met the inclusion criteria. These included one patient with painful posttraumatic arthritis, two patients with failed hemiarthroplasty, and one patient with a failed Sauvé–Kapandji procedure. Mean age at the time of surgery was 42.5 years (range: 35 to 51 years) and mean follow-up was 46 months (range: 36 to 66 months). Preoperative and postoperative assessment included range of motion, grip strength, visual analog pain scale, patient satisfaction, and radiographic evaluation. There was a successful replacement of the DRUJ in all four patients. Final range of motion showed mean pronation of 80 degrees (range: 60 to 90 degrees) and mean supination of 64 degrees (range: 45 to 90 degrees). Final grip strength on the operated extremity was 25.5 kg and averaged 73% of contralateral side. This was an improvement from preoperative grip strength of 14.5 kg visual analog pain scale decreased from 8 to 2.5 following surgery (scale: 1 to 10). Patient satisfaction was 100%. One patient returned to manual labor, one patient returned to office work, and two patients remained off work. Postoperative radiographs depict appropriate alignment of the DRUJ, and there have been no cases of subluxation or dislocation. Implant survival has been 100%. Total DRUJ joint replacement with sigmoid notch resurfacing and distal ulna replacement is an alternative to ulnar head resection in cases of DRUJ arthritis. Total DRUJ arthroplasty with this resurfacing design may provide a means of decreasing pain and restoring DRUJ stability and motion following severe trauma, failed hemiarthroplasty, or failed Sauvé–Kapandji procedure. More experience is needed with this implant to confirm these initial encouraging results. The level of evidence for this study is IV (therapeutic, case series).Keywords
This publication has 57 references indexed in Scilit:
- Ulnar Head ReplacementTechniques in Hand & Upper Extremity Surgery, 2007
- EclypseTechniques in Hand & Upper Extremity Surgery, 2007
- Wear-Induced Osteolysis and Synovial Swelling in a Patient With a Metal–Polyethylene Wrist ProsthesisThe Journal of Hand Surgery, 2006
- The Load-Bearing Characteristics of the Forearm: Pattern of Axial and Bending Force Transmitted Through Ulna and RadiusJournal of Hand Surgery (European Volume), 2006
- The dorsopalmar stability of the distal radioulnar jointThe Journal of Hand Surgery, 2000
- Results of the Modified Sauvé-Kapandji Procedure in the Treatment of Chronic Posttraumatic Derangement of the Distal Radioulnar Joint*Journal of Bone and Joint Surgery, 1998
- Dynamic radio-ulnar convergence after the Darrach procedure.1996
- Matched distal ulnar resectionThe Journal of Hand Surgery, 1986
- Ulnar impingement syndrome.1985
- Implant Arthroplasty for Disabilities of the Distal Radioulnar Joint Use of a Silicone Rubber Capping Implant Following Resection of the Ulnar HeadOrthopedic Clinics of North America, 1973