Failure to Deliver Therapy by a Riata Lead with Internal Wire Externalization and Normal Electrical Parameters During Routine Interrogation
- 21 May 2012
- journal article
- case report
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 24 (1) , 94-96
Abstract
Failure to Deliver Therapy by a Riata Lead. We present a case of failure to deliver a shock by a St. Jude Medical defibrillator involving a Riata lead that was discovered incidentally while the device was attempting to deliver inappropriate therapy. Routine interrogation, including high voltage (HV) impedance, failed to reveal any abnormality. Failure to deliver therapy was confirmed during DFT testing, which revealed extremely low HV impedance only while attempting to deliver therapy. Fluoroscopy indicated moderate externalization of internal wires. This case highlights an under‐recognized issue with St. Jude Medical systems, namely the possibility that therapy may not be delivered despite the presence of normal electrical parameters during routine surveillance. (J Cardiovasc Electrophysiol, Vol. 24, pp. 94‐96, January 2013)Keywords
This publication has 6 references indexed in Scilit:
- Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter–defibrillator leadsInternational Journal of Cardiology, 2013
- Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defectHeart Rhythm, 2012
- Unrecognized Failure of a Narrow Caliber Defibrillation Lead: The Role of Defibrillation Threshold Testing in Identifying an Unprotected IndividualPacing and Clinical Electrophysiology, 2012
- Insulation Defects of Thin High-Voltage ICD Leads:Journal of Cardiovascular Electrophysiology, 2011
- Premature Failure of a Riata Defibrillator Lead Without Impedance Change or Inappropriate Sensing: A Case Report and Review of the LiteratureJournal of Cardiovascular Electrophysiology, 2011
- Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient PopulationJournal of Cardiovascular Electrophysiology, 2010