Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation
Authors:
- Astrid N L HermansAHAstrid N L Hermans
- Jonas L IsaksenJIJonas L Isaksen
- Monika GawalkoMGMonika Gawalko
- Nikki A H A PluymaekersNPNikki A H A Pluymaekers
- Rachel M J van der VeldenRVRachel M J van der Velden
- Justin G L M LuermansJLJustin G L M Luermans
- Martin ManningerMMMartin Manninger
- Joost LumensJLJoost Lumens
- Jørgen K KantersJKJørgen K Kanters
Open Access
- 7 February 2023
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 25 (3) , 835-844
Abstract
Although mobile health tools using photoplethysmography (PPG) technology have been validated for the detection of atrial fibrillation (AF), their utility for heart rate assessment during AF remains unclear. Therefore, we aimed to evaluate the accuracy of continuous PPG-based 1 min mean heart rate assessment during AF. Persistent AF patients were provided with Holter electrocardiography (ECG) (for ≥24 h) simultaneously with a PPG-equipped smartwatch. Both the PPG-based smartwatch and Holter ECG automatically and continuously monitored patients’ heart rate/rhythm. ECG and PPG recordings were synchronized and divided into 1 min segments, from which a PPG-based and an ECG-based average heart rate estimation were extracted. In total, 47 661 simultaneous ECG and PPG 1 min heart rate segments were analysed in 50 patients (34% women, age 73 ± 8 years). The agreement between ECG-determined and PPG-determined 1 min mean heart rate was high [root mean squared error (RMSE): 4.7 bpm]. The 1 min mean heart rate estimated using PPG was accurate within ±10% in 93.7% of the corresponding ECG-derived 1 min mean heart rate segments. PPG-based 1 min mean heart rate estimation was more often accurate during night-time (97%) than day-time (91%, P < 0.001) and during low levels (96%) compared to high levels of motion (92%, P < 0.001). A neural network with a 10 min history of the recording did not further improve the PPG-based 1 min mean heart rate assessment [RMSE: 4.4 (95% confidence interval: 3.5–5.2 bpm)]. Only chronic heart failure was associated with a lower agreement between ECG-derived and PPG-derived 1 min mean heart rates (P = 0.040). During persistent AF, continuous PPG-based 1 min mean heart rate assessment is feasible in 60% of the analysed period and shows high accuracy compared with Holter ECG for heart rates <110 bpm.Keywords
This publication has 24 references indexed in Scilit:
- The Current State of Mobile Phone Apps for Monitoring Heart Rate, Heart Rate Variability, and Atrial Fibrillation: Narrative ReviewJMIR mHealth and uHealth, 2019
- The validation of smartphone applications for heart rate measurementAnnals of Medicine, 2018
- Smart watches for heart rate assessment in atrial arrhythmiasInternational Journal of Cardiology, 2018
- Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillationPhysiological Measurement, 2018
- Rate control in atrial fibrillationThe Lancet, 2016
- Heart rate dynamics distinguish among atrial fibrillation, normal sinus rhythm and sinus rhythm with frequent ectopyPhysiological Measurement, 2015
- Lenient versus Strict Rate Control in Patients with Atrial FibrillationNew England Journal of Medicine, 2010
- A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjectsJournal of Medical Engineering & Technology, 2009
- Rhythm or rate control in atrial fibrillation—Pharmacological Intervention in Atrial Fibrillation (PIAF): a randomised trialThe Lancet, 2000
- Long Short-Term MemoryNeural Computation, 1997