Stainless steel wear debris of a scoliotic growth guidance system has little local and systemic effect in an animal model
Authors:
- 13 February 2018
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 36 (7) , 1980-1990
Abstract
Options to treat early-onset scoliosis include guided-growth systems with sliding action between rods and pedicle screws. The wear was previously measured in an in vitro test, and in this in vivo rabbit model, we evaluated the local and systemic biological response to the stainless steel debris. Compared to the previous study, a relatively higher volume of representative wear particles with a median particle size of 0.84 µm were generated. Bolus dosages were injected into the epidural space at L4-L5 for a minimum of 36 rabbits across three treatment groups (negative control, 1.5 mg, and 4.0 mg) and two timepoints (12 and 24 weeks). Gross pathology evaluated distant organs and the injection site with a dorsal laminectomy to examine the epidural space and dosing site. Peri-implanted particle tissues were stained for immunohistochemical and quantitatively analyzed for IL-6 and TNF-α cytokines. Based on ISO 10993-6:2007 scoring, particles in the high-dose group were primarily non-irritant (12 weeks) with one slightly irritant. At 24 weeks, inflammatory cell infiltration was nonexistent to minimal with all groups considered non-irritant at the injection site. Material characterization confirmed that particles detected in distant organs were stainless steel or contaminants. At 12 weeks, stainless steel groups demonstrated statistically increased amounts of cytokine levels compared to control but there was a statistical decrease for both at 24 weeks. These findings indicate that stainless steel wear debris, comparable to the expected usage from a simulated growth guidance system, had no discernible untoward biological effects locally and systemically in an animal model. This article is protected by copyright. All rights reservedKeywords
Funding Information
- Medtronic, Inc.
This publication has 31 references indexed in Scilit:
- Cobalt-Alloy Implant Debris Induce HIF-1α Hypoxia Associated Responses: A Mechanism for Metal-Specific Orthopedic Implant FailurePLOS ONE, 2013
- Surgical Technique: Modern Luqué Trolley, a Self-growing Rod TechniqueClinical Orthopaedics and Related Research, 2011
- Inflammatory response against different carbon fiber-reinforced PEEK wear particles compared with UHMWPE in vivoActa Biomaterialia, 2010
- Shilla Growing Rods in a Caprine Animal Model: A Pilot StudyClinical Orthopaedics and Related Research, 2010
- The spinal cord dura mater reaction to nitinol and titanium alloy particles: a 1-year study in rabbitsEuropean Spine Journal, 2007
- Wear and Corrosion in Retrieved Thoracolumbar Posterior Internal FixationSpine, 2006
- Evaluation of the biological response of wear debrisThe Spine Journal, 2004
- The effect of spinal instrumentation particulate wear debrisThe Spine Journal, 2003
- The Effect of Titanium Particulate on Development and Maintenance of a Posterolateral Spinal ArthrodesisSpine, 2002
- Long-Term Follow-Up of Patients with Untreated Scoliosis A Study of Mortality, Causes of Death, and SymptomsSpine, 1992