An improved method for the removal of ring artifacts in high resolution CT imaging
Authors:
Open Access
- 27 April 2012
- journal article
- research article
- Published by Springer Nature in EURASIP Journal on Advances in Signal Processing
- Vol. 2012 (1) , 93
Abstract
In high resolution computed tomography (CT) using flat panel detectors, imperfect or defected detector elements cause stripe artifacts in sinogram which results in concentric ring artifacts in the image. Such ring artifacts obscure image details in the regions of interest of the image. In this article, novel techniques are proposed for the detection, classification, and correction of ring artifacts in the sinogram domain. The proposed method is suitable for multislice CT with parallel or fan beam geometry. It can also be employed for ring artifact removal in 3D cone beam volume CT by adopting a sinogram by sinogram processing technique. The detection algorithm is based on applying data driven thresholds on the mean curve and difference curve of the sinogram. The ring artifacts are classified into three types and a separate correction algorithm is used for each class. The performance of the proposed techniques is evaluated on a number of real micro-CT images. Experimental results corroborate that the proposed algorithm can remove ring artifacts from micro-CT images more effectively as compared to other recently reported techniques in the literature.Keywords
This publication has 18 references indexed in Scilit:
- Removal of ring artifacts in CT imaging through detection and correction of stripes in the sinogramPhysics in Medicine & Biology, 2010
- A Self-Adaptive Approach for the Detection and Correction of Stripes in the Sinogram: Suppression of Ring Artifacts in CT ImagingEURASIP Journal on Advances in Signal Processing, 2010
- Removal of ring artifacts in computed tomographic imaging using iterative center weighted median filterComputers in Biology and Medicine, 2010
- Ring artifact correction for high-resolution micro CTPhysics in Medicine & Biology, 2009
- Comparison of ring artifact correction methods for flat-detector CTPhysics in Medicine & Biology, 2009
- Stripe and ring artifact removal with combined wavelet—Fourier filteringOptics Express, 2009
- Image quality assessment: from error visibility to structural similarityIEEE Transactions on Image Processing, 2004
- Minimax estimation via wavelet shrinkageThe Annals of Statistics, 1998
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984
- Smoothing and Differentiation of Data by Simplified Least Squares Procedures.Analytical Chemistry, 1964