Evaluation of the detection of blasts by Sysmex hematology instruments, CellaVision DM96, and manual microscopy using flow cytometry as the confirmatory method
Authors:
- 11 February 2019
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 41 (3) , 338-344
Abstract
Introduction: The aim of the present study was to evaluate the diagnostic ability of blast flags generated by Sysmex instruments (XE/XN) by comparing with immunophenotyping by flow cytometry (IFCM). Additionally, the ability of manual microscopy and CellaVision DM96 (pre‐ and reclassification) to predict the presence of “true” blasts was investigated.Methods: Blood samples (n = 240) with suspect pathology flags reported by the XE were collected from the daily workload and examined by the XN, by manual microscopy, by CellaVision DM96 and by IFCM (CytoDiff Panel).Results: The ROC analysis for blasts showed an area under the curve of 0.64 (“Blasts?”) (XE), 0.57 (“Blasts/Abn Lympho?”) (XN), 0.75 (CellaVision preclassification procedure), 0.78 (CellaVision reclassification procedure), and 0.81 (manual microscopy). The sensitivity of blast detection varied between the methods from 0.41 (XE) to 0.90 (XN), and the specificity varied from 0.17 (XN) to 0.95 (CellaVision reclassification).Conclusions: The CellaVision reclassification procedure has a diagnostic ability for predicting blasts close to that of manual microscopy. The blood smear methods show a notable number of false negative results. The Sysmex XN reported a higher rate of true positive blast flags than the XE. Taken together, the CytoDiff method could be a useful alternative to smear examination to correctly identify blasts.Keywords
This publication has 21 references indexed in Scilit:
- Performance evaluation of the new hematology analyzer Sysmex XN‐seriesInternational Journal of Laboratory Hematology, 2014
- ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte countingInternational Journal of Laboratory Hematology, 2014
- The Usefulness of Blast Flags on the Sysmex XE-5000 Is QuestionableAmerican Journal of Clinical Pathology, 2013
- Toward a reference method for leukocyte differential counts in blood: Comparison of three flow cytometric candidate methodsCytometry Part A, 2012
- Improved Flagging Rates on the Sysmex XE-5000 Compared With the XE-2100 Reduce the Number of Manual Film Reviews and Increase Laboratory ProductivityAmerican Journal of Clinical Pathology, 2011
- Reliable, Accurate Determination of the Leukocyte Differential of Leukopenic Samples by Using Hematoflow MethodAnnals of Laboratory Medicine, 2011
- Refining the white blood cell differential: The first flow cytometry routine applicationCytometry Part A, 2010
- Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis systemInternational Journal of Laboratory Hematology, 2009
- “6 markers/5 colors” extended white blood cell differential by flow cytometryCytometry Part A, 2007
- Examination of peripheral blood films using automated microscopy; evaluation of Diffmaster Octavia and Cellavision DM96Journal of Clinical Pathology, 2006