Simulation analysis of a wearable dry EEG electrodes for epilepsy monitoring
- 1 December 2017
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
Wearable electroencephalography (EEG) system can be worn for prolonged periods and have been widely used to reduce susceptibility to artifacts. It can improve localization of epileptic foci, and help doctors to optimize the treatment for patients to control epileptic seizures effectively. Commercial metal-based dry EEG sensors are capable of providing signal acquisition, however it is uncomfortable for patients since the sensor's material property is too hard to wear. In this study, design and fabrication process of dry EEG sensor are proposed for epilepsy monitoring. A medical grade silicone, Polydimethylsiloxane (PDMS), was molded as deformable substrate of the sensor. Virtual prototyping was created and simulated by ANSYS mechanical structural module. Force measurement of the fabricated sensor was conducted and discussed. The actual and simulated of sensor's stiffness were characterized and discussed.Keywords
This publication has 4 references indexed in Scilit:
- A Flexible and Wearable EEG Device2016
- Usability of four commercially-oriented EEG systemsJournal of Neural Engineering, 2014
- Mechanical characterization of bulk Sylgard 184 for microfluidics and microengineeringJournal of Micromechanics and Microengineering, 2014
- Evaluation of commercially available electrodes and gels for recording of slow EEG potentialsClinical Neurophysiology, 2005