High‐resolution ultrashort echo time (UTE) imaging on human knee with AWSOS sequence at 3.0 T

Authors:

Abstract
Purpose: To demonstrate the technical feasibility of high‐resolution (0.28–0.14 mm) ultrashort echo time (UTE) imaging on human knee at 3T with the acquisition‐weighted stack of spirals (AWSOS) sequence.Materials and Methods: Nine human subjects were scanned on a 3T MRI scanner with an 8‐channel knee coil using the AWSOS sequence and isocenter positioning plus manual shimming.Results: High‐resolution UTE images were obtained on the subject knees at TE = 0.6 msec with total acquisition time of 5.12 minutes for 60 slices at an in‐plane resolution of 0.28 mm and 10.24 minutes for 40 slices at an in‐plane resolution of 0.14 mm. Isocenter positioning, manual shimming, and the 8‐channel array coil helped minimize image distortion and achieve high signal‐to‐noise ratio (SNR).Conclusion: It is technically feasible on a clinical 3T MRI scanner to perform UTE imaging on human knee at very high spatial resolutions (0.28–0.14 mm) within reasonable scan time (5–10 min) using the AWSOS sequence. J. Magn. Reson. Imaging 2012;35:204‐210.