Soft Tissue Balance Profiles Differ Between Manual and Robotically Assessed Gaps in Total Knee Arthroplasty
- 29 November 2025
- journal article
- Published by Wiley in The International Journal of Medical Robotics and Computer Assisted Surgery
- Vol. 21 (6) , e70124
Abstract
Background: This study examines the agreement and error between manual gap assessment, and assessment with a robotic soft‐tissue tensioning device in total knee arthroplasty across multiple surgeons. Methods: 517 consecutive robotic‐assisted TKAs were retrospectively reviewed across six surgeons following ethics approval. Robotic and manual mediolateral (ML) balance and medial and lateral gaps were compared using Intraclass Correlation Coefficient (ICC). ΔML balance (manual‐robotic) was compared between right and left knees. Results: Manual and robotic ML balance showed poor agreement (ICC = 0.368). Surgeons had low to moderate correlations (ICC range: 0.152–0.633) and significant differences throughout flexion. Four measured greater medial gaps manually and two measured greater lateral gaps compared to robotics. All showed significant ΔML balance differences between the leg sides ( p < 0.05), with surgeon‐specific variations in gap patterns. Conclusions: Manual ML balance correlated poorly and differed significantly from robotic assessment, with surgeon‐specific variation. Surgeons should appreciate the variability in subjective versus objective gap measurements.Keywords
This publication has 10 references indexed in Scilit:
- Navigated instrumentation and ligament tensioning device enhances initial gap acquisition during total knee arthroplasty procedure: A cadaveric studyJournal of Experimental Orthopaedics, 2025
- Surgeon-Applied Stress and a Ligament Tensor Instrument Provide a Similar Assessment of Preresection Flexion Laxity During Robotic Total Knee ArthroplastyArthroplasty Today, 2024
- Effects of Surgeon Handedness on the Outcomes of Unicompartmental Knee Arthroplasty: A Single Center's ExperienceOrthopaedic Surgery, 2022
- Arthroplasty Surgeons Differ in Their Intraoperative Soft Tissue Assessments: A Study in Human Cadavers to Quantify Surgical Decision-making in TKAClinical Orthopaedics and Related Research, 2022
- Inter-Rater Reliability of Clinical Testing for Laxity After Knee ArthroplastyThe Journal of Arthroplasty, 2022
- Imageless, robotic-assisted total knee arthroplasty combined with a robotic tensioning system can help predict and achieve accurate postoperative ligament balanceArthroplasty Today, 2019
- How Accurately Can Soft Tissue Balance Be Determined in Total Knee Arthroplasty?The Journal of Arthroplasty, 2019
- Laxity Profiles in the Native and Replaced Knee—Application to Robotic-Assisted Gap-Balancing Total Knee ArthroplastyThe Journal of Arthroplasty, 2018
- Can We Really “Feel” a Balanced Total Knee Arthroplasty?The Journal of Arthroplasty, 2016
- Accuracy of soft tissue balancing in total knee arthroplastyThe Journal of Arthroplasty, 2000