FLUCTUATIONS ON THE HUMPHREY AND OCTOPUS PERIMETERS

Authors:
  • RBRS BRENTON
    RS BRENTON
  • WAWA ARGUS
    WA ARGUS

  • 1 May 1987
    • journal article
    • research article
    • Vol. 28  (5) , 767-771
    • PMID:3570687
Abstract
Fluctuation of threshold determinations must be considered when evaluating visual field abnormalities. The short- and long-term threshold fluctuation were measured on 36 normal subjects using the Octopus and Humphrey automated perimeters. Subjects were tested using similar threshold programs that tested 16 locations twice. Short-term fluctuation was greater on the Octopus perimeter (1.6 .+-. 0.3 dB) than on the Humphrey perimeter (1.3 .+-. 0.3 dB; P < 0.001). The heterogenous long-term fluctuation was greater on the Octopus perimeter (1.6 .+-. 0.4 dB) than on the Humphrey perimeter (1.3 .+-. 0.5 dB; P < 0.05). The homogenous long-term fluctuation was near 0 for both machines. There are numerous variables that can alter the reproducibility of perimetry. Reproducibility can be estimated by short- and long-term fluctuation, but differences between the Octopus and Humphrey also influence their fluctuation.