Comparison meta-analysis of two swept-source optical coherence biometers
Authors:
Open Access
- 29 December 2025
- journal article
- research article
- Published by Wolters Kluwer Health in Indian Journal Of Ophthalmology
- Vol. 74 (1) , 23-28
Abstract
This systematic review and meta-analysis aims to evaluate the comparability of two swept-source optical coherence tomography (SS-OCT) biometers, Anterion and IOLMaster 700. This systematic review and meta-analysis reviewed articles published until October 1, 2023. Eligibility criteria included a comparative study design, absence of ocular diseases except cataract or refractive errors, same day measurements with both devices, and outcomes such as axial length (AL), mean keratometry, astigmatism vectors, anterior chamber depth (ACD), central corneal thickness, lens thickness (LT), corneal diameter, white-to-white (WTW), and intraocular lens (IOL) power for emmetropia. This systematic review incorporated 17 studies (2943 eyes). The mean difference of AL measurements obtained with Anterion and IOLMaster 700 was -0.01 [-0.09; 0.06]. Anterion provided a significantly larger ACD (0.10 mm [0.08-0.12 mm]) and LT (0.04 mm [0.01; 0.07]), while IOLMaster 700 had significantly larger WTW measurements (0.20 mm [-0.24; -0.16]). Corneal measurements tend to be flatter for Anterion, although no significant differences were found between devices. In conclusion, this meta-analysis showed statistically significant differences in WTW, LT, and ACD measurements between devices. AL measurements were comparable between biometers. Anterior mean and steep and flat keratometry tend to be higher with IOLMaster700. These differences may result in a higher IOL power for emmetropia when data are entered from Anterion measurements. Future studies will be needed to establish the influence of these differences on the refractive predictability of IOL power calculation.Keywords
This publication has 30 references indexed in Scilit:
- Comparison of 2 modern swept-source optical biometers—IOLMaster 700 and AnterionGraefe's Archive for Clinical and Experimental Ophthalmology, 2022
- Repetibilidad y comparabilidad de un nuevo tomógrafo con tecnología SS-OCT en biometría ópticaArchivos de la Sociedad Española de Oftalmología, 2022
- Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometersGraefe's Archive for Clinical and Experimental Ophthalmology, 2022
- Agreement of predicted intraocular lens power using swept-source optical coherence tomography and partial coherence interferometryExpert Review of Medical Devices, 2021
- Intraobserver Repeatability of Tomographic, Pachymetric, and Anatomical Measurements in Healthy Eyes Using a New Swept-Source Optical Coherence TopographerCornea, 2021
- Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometerJournal of Cataract & Refractive Surgery, 2021
- Agreement between intraoperative anterior segment spectral-domain OCT and 2 swept-source OCT biometersExpert Review of Medical Devices, 2021
- Comparison of 2 swept-source optical coherence tomography–based biometry devicesJournal of Cataract & Refractive Surgery, 2020
- Review and recommendations for univariate statistical analysis of spherical equivalent prediction error for IOL power calculationsJournal of Cataract & Refractive Surgery, 2020
- Pursuing perfection in intraocular lens calculations: III. Criteria for analyzing outcomesJournal of Cataract & Refractive Surgery, 2017