Abstract
Object: A number of imaging techniques have been introduced to minimize the risk of pedicle screw placement. Intraoperative CT has been recently introduced to assist in spinal instrumentation. The aim of this study was to study the effectiveness of intraoperative CT in enhancing the safety and accuracy of pedicle screw placement. Methods: The authors included all cases from December 2009 through July 2012 in which intraoperative CT scanning was used to confirm pedicle screw placement. Results: A total of 203 patients met the inclusion criteria. Of 1148 screws, 103 screws (8.97%) were revised intraoperatively in 72 patients (35.5%): 14 (18.42%) were revised in the cervical spine (C-2 or C-7), 25 (7.25%) in the thoracic spine, and 64 (8.80%) in the lumbar spine. Compared with screws in the thoracic and lumbar regions, pedicle screws placed in the cervical region were statistically more likely to be revised (p = 0.0061). Two patients (0.99%) required reoperations due to undetected misplacement of pedicle screws. Conclusions: The authors describe one of the first North American experiences using intraoperative CT scanning to confirm the placement of pedicle screws. Compared with a similar cohort of patients from their institution who had pedicle screws inserted via the free-hand technique with postoperative CT, the authors found that the intraoperative CT lowers the threshold for pedicle screw revision, resulting in a statistically higher rate of screw revision in the thoracic and lumbar spine (p < 0.0001). During their 2.5-year experience with the intraoperative CT, the authors did not find a reduction in rates of reoperation for misplaced pedicle screws.Keywords
This publication has 42 references indexed in Scilit:
- Pedicle screw insertion: computed tomography versus fluoroscopic image guidanceInternational Orthopaedics, 2007
- Image-guided spinal surgeryThe International Journal of Medical Robotics and Computer Assisted Surgery, 2006
- Intraoperative Computerized Tomography for Improved Accuracy of Spinal Navigation in Pedicle Screw Placement of the Thoracic Spine2003
- Image-Guided Spine SurgerySpine, 2002
- Präzisionsanalyse navigationsgestützt implantierter thorakaler und lumbaler PedikelschraubenDer Unfallchirurg, 2001
- Transpedicular Screw PlacementSpine, 2001
- Computer-assisted Fluoroscopic Targeting System for Pedicle Screw InsertionNeurosurgery, 2000
- Comparative Results Between Conventional and Computer-Assisted Pedicle Screw Installation in the Thoracic, Lumbar, and Sacral SpineSpine, 2000
- A Comparison of Fluoroscopy and Computed Tomography-derived Volumetric Multiple Exposure Transmission Holography for the Guidance of Lumbar Pedicle Screw InsertionNeurosurgery, 1995
- Accuracy of Pedicle Screw Placement in Lumbar Fusions by Plain Radiographs and Computed TomographySpine, 1995