Comparison of Prospectively Generated Glioma Treatment Plans Clinically Delivered on Magnetic Resonance Imaging (MRI)-Linear Accelerator (MR-Linac) Versus Conventional Linac: Predicted and Measured Skin Dose
Open Access
- 7 September 2022
- journal article
- research article
- Published by SAGE Publications in Technology in Cancer Research & Treatment
- Vol. 21
Abstract
Introduction: Magnetic resonance imaging-linear accelerator radiotherapy is an innovative technology that requires special consideration for secondary electron interactions within the magnetic field, which can alter dose deposition at air–tissue interfaces. As part of ongoing quality assurance and quality improvement of new radiotherapy technologies, the purpose of this study was to evaluate skin dose modelled from the treatment planning systems of a magnetic resonance imaging-linear accelerator and a conventional linear accelerator, and then correlate with in vivo measurements of delivered skin dose from each linear accelerator. Methods: In this prospective cohort study, 37 consecutive glioma patients had treatment planning completed and approved prior to radiotherapy initiation using commercial treatment planning systems: a Monte Carlo-based algorithm for magnetic resonance imaging-linear accelerator or a convolution-based algorithm for conventional linear accelerator. In vivo skin dose was measured using an optically stimulated luminescent dosimeter. Results: Monte Carlo-based magnetic resonance imaging-linear accelerator plans and convolution-based conventional linear accelerator plans had similar dosimetric parameters for target volumes and organs-at-risk. However, magnetic resonance imaging-linear accelerator plans had 1.52 Gy higher mean dose to air cavities ( P < .0001) and 1.10 Gy higher mean dose to skin ( P < .0001). In vivo skin dose was 14.5% greater for magnetic resonance imaging-linear accelerator treatments ( P = .0027), and was more accurately predicted by Monte Carlo-based calculation ( ρ = 0.95, P < .0001) versus convolution-based ( ρ = 0.80, P = .0096). Conclusion: This is the first prospective dosimetric comparison of glioma patients clinically treated on both magnetic resonance imaging-linear accelerator and conventional linear accelerator. Our findings suggest that skin doses were significantly greater with magnetic resonance imaging-linear accelerator plans but correlated better with in vivo measurements of actual skin dose from delivered treatments. Future magnetic resonance imaging-linear accelerator planning processes are being designed to account for skin dosimetry and treatment delivery.Keywords
This publication has 35 references indexed in Scilit:
- Evaluation of a commercial MRI Linac based Monte Carlo dose calculation algorithm with geant 4Medical Physics, 2016
- Compensating for the impact of non-stationary spherical air cavities on IMRT dose delivery in transverse magnetic fieldsPhysics in Medicine & Biology, 2015
- Evaluation of volumetric modulated arc therapy for cranial radiosurgery using multiple noncoplanar arcsMedical Physics, 2011
- GPUMCD: A new GPU‐oriented Monte Carlo dose calculation platformMedical Physics, 2011
- Simultaneous multi-modality ROI delineation in clinical practiceComputer Methods and Programs in Biomedicine, 2009
- Dose optimization for the MRI-accelerator: IMRT in the presence of a magnetic fieldPhysics in Medicine & Biology, 2007
- The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) Statement: Guidelines for Reporting Observational StudiesAnnals of Internal Medicine, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: impact of the surface orientation on the entrance and exit dose due to the transverse magnetic fieldPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- Radiotherapy plus Concomitant and Adjuvant Temozolomide for GlioblastomaThe New England journal of medicine, 2005