Amplitude demodulation of the electrocardiogram signal (ECG) for respiration monitoring and compensation during MR examinations
- 1 July 1997
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 38 (1) , 129-136
Abstract
A method for respiration monitoring during MR‐sequences using the electrocardiogram signal (ECG) is presented. The basic principle is known in conventional patient monitoring and is based on the fact that the amplitude of the ECG‐signal varies with respiratory motion. A demodulation of this “cardio‐respiratory signal” yields a respiration curve that can be used for patient monitoring, triggering, or respiration compensation of MRI sequences. Since no drift is introduced by digital demodulation, this method is superior to analog signal processing. The proposed method was applied during MR examinations using an optically coupled ECG sensor that has the advantage of almost negligible interference due to the MR sequences (r= 0.79 to 0.98) with the position of the diaphragm in inferior‐superior direction measured during breath‐holding MRI.Keywords
This publication has 20 references indexed in Scilit:
- Synchronization device for electrocardiography-gated echo-planar imaging.Radiology, 1995
- Electrocardiogram acquisition during MR examinations for patient monitoring and sequence triggeringMagnetic Resonance in Medicine, 1994
- Patient monitoring during clinical MR imaging.Radiology, 1992
- Burns Associated with Pulse Oximetry during Magnetic Resonance ImagingAnesthesiology, 1991
- Respiratory gating in magnetic resonance imaging: Improved image quality over non-gated images for equal scan timeClinical Imaging, 1991
- The effect of respiration-induced heart movements on the ECGIEEE Transactions on Biomedical Engineering, 1989
- Software QRS detection in ambulatory monitoring — a reviewMedical & Biological Engineering & Computing, 1984
- Optimal QRS detectorMedical & Biological Engineering & Computing, 1983
- QRS wave detectionMedical & Biological Engineering & Computing, 1980
- Influence of respiration on recording cardiac potentialsThe American Journal of Cardiology, 1967