Angle-fixated Anterior Chamber Phakic Intraocular Lens for Myopia of -7 to -19 Diopters
Authors:
- Jean L ArneJAJean L Arne
- Yves BokobzaYBYves Bokobza
- Joseph ColinJCJoseph Colin
- Jean L GeorgeJGJean L George
- Francoise LagoutteFLFrancoise Lagoutte
- Pierre LesurePLPierre Lesure
- Michel MontardMMMichel Montard
- Jean J SaragoussiJSJean J Saragoussi
- Patrick SecheyronPSPatrick Secheyron
- 1 May 1998
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 14 (3) , 282-293
Abstract
PURPOSE: To assess the safety and efficacy of a second generation, anterior chamber phakic intraocular lens (Baikoff Model ZB5M) in patients with high myopia. METHODS: One hundred and twenty-one patients (134 eyes) with myopia of -7.00 to -18.80 D were implanted with the ZB5M lens and followed for 18 to 52 months; the number of eyes evaluated was: 6 months (104 eyes), 1 year (91 eyes), 18 months (78 eyes), 2 years (68 eyes), and 3 years (35 eyes). RESULTS: Postoperative spherical equivalent refraction averaged -1.00 D and the error in refractive correction (achieved minus intended) averaged -0.40 D during the first 2 years, increasing -to -1.30 D refraction and -0.60 D error in refractive correction at 3 years. At 2 years, approximately 40% of eyes had a spherical equivalent refraction within ±0.50 D, and 65% within ±1.00 D. The uncorrected distance visual acuity was 0.048 at baseline and 0.5 at 3 years; near visual acuity was 0.21 at baseline and approximately 0.7 over the 3 years of follow-up. Spectacle-corrected distance visual acuity at baseline was 0.54; it improved to 0.7. Near visual acuity was 0.65 at baseline and it improved to approximately 0.75. Endothelial cell counts in the central and peripheral cornea were reduced by an average 3.3% at 6 months, declining an additional 1% to 2% over the remaining follow-up period. Regression analyses indicated that most of the endothelial cell loss was due to surgery. Additional complications included halos/glare in 37 of 133 eyes (27.8%) and iris retraction with pupillary ovalization in 30 of 133 eyes (22.6%). The intraocular lens was exchanged in four of 133 eyes (3.0%) and removed in three of 133 eyes (2.3%), the latter because of halos (one eye) and a flat anterior chamber with severe inflammation (two eyes). CONCLUSIONS: Implantation of the Baikoff ZB5M lens in the anterior chamber of phakic eyes significantly reduced high myopia and produced a stable refractive outcome over the 3 years, accompanied by marked improvement in uncorrected distance visual acuity and minimal, non-progressive damage to the corneal endothelium. Frequent complications included pupillary ovalization and halos/glare. Improvements in accuracy of IOL power calculations are needed. [J Refract Surg 1998;14:282-293]Keywords
This publication has 11 references indexed in Scilit:
- Endothelial changes in phakic eyes with anterior chamber intraocular lenses to correct high myopiaJournal of Cataract & Refractive Surgery, 1996
- Retinal detachment following anterior chamber lens implantation for the correction of ultra-high myopia in phakic eyes.British Journal of Ophthalmology, 1993
- Retinal Detachment as a Potential Hazard in Surgical Correction of Severe Myopia With Phakic Anterior Chamber LensesAmerican Journal of Ophthalmology, 1993
- Damage to the Corneal Endothelium by Minus Power Anterior Chamber Intraocular LensesJournal of Refractive Surgery, 1991
- Damage to the Corneal Endothelium From Anterior Chamber Intraocular Lenses in Phakic Myopic EyesJournal of Refractive Surgery, 1991
- Phakic Anterior Chamber Intraocular LensesInternational Ophthalmology Clinics, 1991
- Some Optical Aspects of Implantation of an IOL in a Myopic EyeEuropean journal of Implant and Refractive Surgery, 1989
- Changes in the normal corneal endothelial cellular pattern as a function of ageCurrent Eye Research, 1985
- Effect of age on the endothelial cell count in the normal eye.British Journal of Ophthalmology, 1982
- Specular Microscopy of Human Corneal Endothelium in VIVOAmerican Journal of Ophthalmology, 1976