Application of refocused steady‐state free‐precession methods at 1.5 and 3 T to in vivo high‐resolution MRI of trabecular bone: Simulations and experiments

Authors:

Abstract
Purpose: To evaluate the potential of fully‐balanced steady‐state free‐precession (SSFP) sequences in in vivo high‐resolution (HR) MRI of trabecular bone at field strengths of 1.5 and 3 T by simulation and experimental methods.Materials and Methods: Using simulation studies, refocused SSFP acquisition was optimized for our imaging purposes with a focus on signal‐to‐noise ratio (SNR) and SNR efficiency. The signal behavior in trabecular bone was estimated using a magnetostatic model of the trabecular bone and marrow. Eight normal volunteers were imaged at the proximal femur, calcaneus, and the distal tibia on a GE Signa scanner at 1.5 and at 3 T with an optimized single‐acquisition SSFP sequence (three‐dimensional FIESTA) and an optimized multiple‐acquisition SSFP sequence (three‐dimensional FIESTA‐c). Images were also acquired with a fast gradient echo (FGRE) sequence for evaluation of the SNR performance of SSFP methods.Results: Refocused SSFP images outperformed FGRE acquisitions in both SNR and SNR efficiency at both field strengths. At 3 T, susceptibility effects were visible in FIESTA and FGRE images and much reduced in FIESTA‐c images. The magnitude of SNR boost at 3 T was closely predicted by simulations.Conclusion: Single‐acquisition SSFP (at 1.5 T) and multiple‐acquisition SSFP (at 3 T) hold great potential for HR‐MRI of trabecular bone. J. Magn. Reson. Imaging 2005;21:818–825.