Multicentre performance evaluation of the E170 Module for MODULAR ANALYTICS
Authors:
- Daniel W. ChanDCDaniel W. Chan
- Lori SokollLSLori Sokoll
- Roland ImdahlRIRoland Imdahl
- Masaji KobayashiMKMasaji Kobayashi
- Erike YamadaEYErike Yamada
- Diana J. LiljeDLDiana J. Lilje
- Hilmar LutheHLHilmar Luthe
- Jochen MeissnerJMJochen Meissner
- Gianni MesseriGMGianni Messeri
- Alessandra CelliACAlessandra Celli
- Paola TozziPTPaola Tozzi
- 1 January 2004
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 42 (10) , 1186-202
Abstract
The E170 module was evaluated at 13 sites in an international multicentre study. The objective of the study was to assess the analytical performance of 49 analytes, and to collect feedback on the system’s reliability and practicability. The typical, within-run coefficients of variation (CVs) for most of the quantitative assays ranged between 1 and 2% while a range of 2–4% was achieved with the infectious disease methods. Total precision CVs were found to be within the manufacturer’s expected performance ranges, demonstrating good concordance of the system’s measuring channels and a high reproducibility during the 2–4-week trial period. The functional sensitivity of 11 selected assays met the clinical requirements (e.g., thyreotroponin (TSH) 0.008 mU/l, troponin T 0.02 µg/l, total prostate-specific antigen (PSA) 0.03 µg/l). The E170 showed no drift during an 8-hour period and no relevant reagent carryover. Accuracy was confirmed by ring trial experiments and method comparisons vs. ElecsysKeywords
This publication has 2 references indexed in Scilit:
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- CAEv–A program for computer aided evaluationJournal of Analytical Methods in Chemistry, 1991