Multicentre performance evaluation of the E170 Module for MODULAR ANALYTICS

Authors:
  • DCDaniel W. Chan
    Daniel W. Chan
  • LSLori Sokoll
    Lori Sokoll
  • RIRoland Imdahl
    Roland Imdahl
  • MKMasaji Kobayashi
    Masaji Kobayashi
  • EYErike Yamada
    Erike Yamada
  • DLDiana J. Lilje
    Diana J. Lilje
  • HLHilmar Luthe
    Hilmar Luthe
  • JMJochen Meissner
    Jochen Meissner
  • GMGianni Messeri
    Gianni Messeri
  • ACAlessandra Celli
    Alessandra Celli
  • PTPaola Tozzi
    Paola Tozzi

Abstract
The E170 module was evaluated at 13 sites in an international multicentre study. The objective of the study was to assess the analytical performance of 49 analytes, and to collect feedback on the system’s reliability and practicability. The typical, within-run coefficients of variation (CVs) for most of the quantitative assays ranged between 1 and 2% while a range of 2–4% was achieved with the infectious disease methods. Total precision CVs were found to be within the manufacturer’s expected performance ranges, demonstrating good concordance of the system’s measuring channels and a high reproducibility during the 2–4-week trial period. The functional sensitivity of 11 selected assays met the clinical requirements (e.g., thyreotroponin (TSH) 0.008 mU/l, troponin T 0.02 µg/l, total prostate-specific antigen (PSA) 0.03 µg/l). The E170 showed no drift during an 8-hour period and no relevant reagent carryover. Accuracy was confirmed by ring trial experiments and method comparisons vs. Elecsys