Comparing New Smartphone-Connected Handheld Ultrasound Device vs. Traditional Ultrasound in Vitreo-Retinal Disease Diagnosis
Authors:
- Giovanni RubegniGRGiovanni Rubegni
- Marco ZeppieriMarco Zeppieri
- Andrea RussoARAndrea Russo
- Niccolò CastellinoNCNiccolò Castellino
- Mario FruschelliMFMario Fruschelli
- Theodora HadjistilianouTHTheodora Hadjistilianou
- Linda TognettiLinda Tognetti
- Matteo OrioneMOMatteo Orione
- Luca LanzafameLLLuca Lanzafame
- Caterina GaglianoCGCaterina Gagliano
- Alessandra CartocciACAlessandra Cartocci
- Gian Marco TosiGian Marco Tosi
Open Access
- 5 September 2024
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 14 (17) , 1961
Abstract
(1) Background: Ocular emergencies account for 1.5–3% of emergency department (ED) visits and require urgent diagnosis to prevent serious complications. Ultrasonography is a crucial, non-invasive diagnostic tool for these conditions but traditionally lacks portability and integration with modern electronic smart devices. The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound device in comparison to a conventional cart-based sonographic machine in patients attending to the ED for vitreo-retinal diseases. (2) Methods: three specialists in ophthalmology, with at least 4-year experience in vitreo-retinal diseases and eye ultrasound, evaluated images of 50 eyes with both portable and traditional ultrasound probes. Each specialist made the diagnosis based on the images captured with both probes and then rated their overall image quality and confidence of diagnosis with a five-point Likert scale. The concordance of diagnosis between the two probes was evaluated. (3) Results: The sample comprised 42 patients. Twenty (40%) healthy eyes and thirty eyes with the following vitreo-retinal interface conditions were examined: 12 retinal detachment (24%), 8 vitreous hemorrhage (16%), and 10 posterior vitreous detachment (20%). The overall accuracy of the two devices appeared to be comparable (70.7% vs. 69.3%). The Butterfly iQ+ probe showed similar sensitivity in retinal detachment diagnosis (91.7% vs. 94.4% of the Accutome B-scan Pro), while it showed poor performance in diagnosing posterior vitreous detachment (sensitivity = 27.2%); (4) Conclusions: The Butterfly iQ+ device demonstrated high sensitivity in the diagnosis of retinal detachment. Significant adjustments are still needed to improve the resolution of the vitreous body.Keywords
This publication has 28 references indexed in Scilit:
- Portable Pocket-Sized Ultrasound Scanner for the Evaluation of Lung Involvement in Coronavirus Disease 2019 PatientsUltrasound in Medicine & Biology, 2021
- Accuracy and performance of a new handheld ultrasound machine with wireless systemScientific Reports, 2019
- Point-of-Care Ultrasonography in the Diagnosis of Retinal Detachment, Vitreous Hemorrhage, and Vitreous Detachment in the Emergency DepartmentJAMA Network Open, 2019
- Point‐of‐Care Ocular Ultrasound for the Diagnosis of Retinal Detachment: A Systematic Review and Meta‐AnalysisAcademic Emergency Medicine, 2019
- Can emergency physicians accurately distinguish retinal detachment from posterior vitreous detachment with point-of-care ocular ultrasound?The American Journal of Emergency Medicine, 2018
- Flashes and FloatersPrimary Care: Clinics in Office Practice, 2015
- The Diagnostic Accuracy of Bedside Ocular Ultrasonography for the Diagnosis of Retinal Detachment: A Systematic Review and Meta-analysisAnnals of Emergency Medicine, 2015
- Use of Ocular Ultrasound for the Evaluation of Retinal DetachmentThe Journal of Emergency Medicine, 2011
- Bedside Ocular Ultrasound for the Detection of Retinal Detachment in the Emergency DepartmentAcademic Emergency Medicine, 2010
- The epidemiology of rhegmatogenous retinal detachment: geographical variation and clinical associationsBritish Journal of Ophthalmology, 2009