Depth dose enhancement of electron beams subject to external uniform longitudinal magnetic fields: A Monte Carlo study
Authors:
- M. A. EarlMEM. A. Earl
- L. MaLML. Ma
- 26 March 2002
- journal article
- Published by Wiley in Medical Physics
- Vol. 29 (4) , 484-491
Abstract
We studied the dose distributions from electron beams subjected to a longitudinal magnetic field applied to them before they reach the phantom. We found that specific combinations of the length and intensity of the magnetic field produced enhancement of the peaks of the central‐axis depth‐dose distributions. TheEGS4Monte Carlo system was used in this study. In the simulations, a uniform axial magnetic field parallel to the electron beam direction was applied to the air gap between the collimation and the phantom. We extensively studied the simplified case of an 18 MeV electron beam point source. Dose deposition was calculated for various magnetic field strengths, distances through which the magnetic field was applied, collimation sizes, and source to collimation distances. The magnetic field strengths varied from 0 to 3 T, the source‐to‐collimation distances studied were 50 and 95 cm, the collimation sizes studied were 10×10 and 20×20 cm2, and the distance through which the field was applied ranged from 10 to 20 cm. Specific combinations of these variables resulted in as much as a 70% enhancement of the peak dose relative to the surface dose. Finally, to determine how the geometry of a real accelerator affects the resulting dose distribution, we performed a full simulation of an Elekta SL20 linear accelerator and compared the results with the ideal case.Keywords
This publication has 9 references indexed in Scilit:
- Reducing loss in lateral charged‐particle equilibrium due to air cavities present in x‐ray irradiated media by using longitudinal magnetic fieldsMedical Physics, 2001
- An apparatus for applying strong longitudinal magnetic fields to clinical photon and electron beamsPhysics in Medicine & Biology, 2001
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995
- The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beamsMedical Physics, 1993
- Electron Transport in $\vec E$ and ${\mathbf{\vec B}}$ Fields1988
- The EGS4 Code System1985
- Enhancement of electron beam dose distributions by longitudinal magnetic fields: Monte Carlo simulations and magnet system optimizationMedical Physics, 1985
- New Developments in High Energy Electron Beam Therapy with the 35 Mev Brown Boveri BetatronRadiology, 1960
- Possible Techniques in Direct-Electron-Beam Tumor TherapyPhysical Review B, 1950