A rare case of Rotablator® driveshaft fracture and successful percutaneous retrieval of a trapped burr using a balloon and GuideLiner®
- 31 August 2016
- journal article
- case report
- Published by Springer Nature in Cardiovascular Intervention and Therapeutics
- Vol. 32 (3) , 294-298
Abstract
A 75-year-old man with a severe calcified atheromatous plaque of the right coronary artery underwent percutaneous coronary intervention using a Rotablator®. During the procedure, the 1.5-mm burr became entrapped, and the driveshaft fractured. While various techniques for retrieving the burr were attempted, a new technique was created. The entrapped burr was sandwiched between an inflated balloon and the GuideLiner®, and a strong back-up force was obtained. Finally, the system was pulled back safely, and the entrapped burr was retrieved without complications. This approach may be useful and safe when coronary flow is preserved despite the trapped burr.Keywords
This publication has 7 references indexed in Scilit:
- Current Status of Rotational AtherectomyJACC: Cardiovascular Interventions, 2014
- Long‐term clinical outcome of rotational atherectomy followed by drug‐eluting stent implantation in complex calcified coronary lesionsCatheterization and Cardiovascular Interventions, 2012
- GuideLiner, a child‐in‐a‐mother catheter for successful retrieval of an entrapped rotablator burrCatheterization and Cardiovascular Interventions, 2011
- Successful retrieval of an entrapped rotablator burr using 5 Fr guiding catheterCatheterization and Cardiovascular Interventions, 2011
- Emergent coronary artery bypass grafting after a broken Rotablator® drive-shaftInteractive CardioVascular and Thoracic Surgery, 2010
- An evidence‐based approach to the use of rotational and directional coronary atherectomy in the era of drug‐eluting stents: When does it make sense?Catheterization and Cardiovascular Interventions, 2008
- Trapped rotablator: Kokesi phenomenonCatheterization and Cardiovascular Interventions, 2000