Evaluation of a Rapid Whole Blood ELISA for Quantification of Troponin I in Patients with Acute Chest Pain

Authors:

Abstract
Background: Troponin I (cTnI) provides important prognostic information in patients with chest pain. We wished to evaluate a rapid, whole-blood analyzer for quantitative point-of-care testing. Methods: A quantitative point-of-care test system (Stratus CS®; Dade-Behring) for cTnI with an incorporated centrifuge was evaluated in 412 patients with chest pain less than 12 h. Results: Results were available within 15 min. CVs were 4.5% at 0.1 μg/L, 4.2% at 0.25 μg/L, and 6.5% at 0.82 μg/L. The detection limit was 0.01 μg/L. The 97.5% percentile in a healthy population was 0.08 μg/L. Based on ROC curve analysis, a threshold of 0.15 μg/L was calculated for the detection of acute myocardial infarction (AMI). With it, sensitivity for the detection of patients with AMI (n = 62) was 63% at arrival and 98% after 4 h (Stratus II®, 48% and 85%, respectively; P P Conclusion: The Stratus CS provided better analytical performance and comparable or better prognostic information than the Stratus II.