Tomographic mammography using a limited number of low‐dose cone‐beam projection images
Authors:
- Alexander StewartASAlexander Stewart
- Martin StantonMSMartin Stanton
- Thomas McCauleyTMThomas McCauley
- Walter PhillipsWPWalter Phillips
- Daniel B. KopansDKDaniel B. Kopans
- Richard H. MooreRMRichard H. Moore
- Jeffrey W. EberhardJEJeffrey W. Eberhard
- Beale Opsahl‐OngBOBeale Opsahl‐Ong
- Loren NiklasonLNLoren Niklason
- Mark B. WilliamsMWMark B. Williams
- 6 February 2003
- journal article
- clinical trial
- Published by Wiley in Medical Physics
- Vol. 30 (3) , 365-380
Abstract
A method is described for using a limited number (typically 10–50) of low‐dose radiographs to reconstruct the three‐dimensional (3D) distribution of x‐ray attenuation in the breast. The method uses x‐ray cone‐beam imaging, an electronic digital detector, and constrained nonlinear iterative computational techniques. Images are reconstructed with high resolution in two dimensions and lower resolution in the third dimension. The 3D distribution of attenuation that is projected into one image in conventional mammography can be separated into many layers (typically 30–80 1‐mm‐thick layers, depending on breast thickness), increasing the conspicuity of features that are often obscured by overlapping structure in a single‐projection view. Schemes that record breast images at nonuniform angular increments, nonuniform image exposure, and nonuniform detector resolution are investigated in order to reduce the total x‐ray exposure necessary to obtain diagnostically useful 3D reconstructions, and to improve the quality of the reconstructed images for a given exposure. The total patient radiation dose can be comparable to that used for a standard two‐view mammogram. The method is illustrated with images from mastectomy specimens, a phantom, and human volunteers. The results show how image quality is affected by various data‐collection protocols.Keywords
Funding Information
- National Institutes of Health
- National Cancer Institute (CA66232, CA69452)
- U.S. Army (DAMD 17‐98‐8329, DAMD 97‐1‐7144)
This publication has 23 references indexed in Scilit:
- Edge-preserving tomographic reconstruction with nonlocal regularizationIEEE Transactions on Medical Imaging, 2002
- Spiral interpolation algorithm for multislice spiral CT. I. TheoryIEEE Transactions on Medical Imaging, 2000
- A projector/backprojector with slice-to-slice blurring for efficient three-dimensional scatter modelingIEEE Transactions on Medical Imaging, 1999
- Globally convergent edge-preserving regularized reconstruction: an application to limited-angle tomographyIEEE Transactions on Image Processing, 1998
- Accelerated iterative transmission CT reconstruction using an ordered subsets convex algorithmIEEE Transactions on Medical Imaging, 1998
- Grouped-coordinate ascent algorithms for penalized-likelihood transmission image reconstructionIEEE Transactions on Medical Imaging, 1997
- Globally convergent algorithms for maximum a posteriori transmission tomographyIEEE Transactions on Image Processing, 1995
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- A generalized EM algorithm for 3-D Bayesian reconstruction from Poisson data using Gibbs priorsIEEE Transactions on Medical Imaging, 1989
- The reconstruction of a three-dimensional structure from projections and its application to electron microscopyProceedings of the Royal Society of London. A. Mathematical and Physical Sciences, 1970