Performance analysis of a film dosimetric quality assurance procedure for IMRT with regard to the employment of quantitative evaluation methods
Authors:
- Peter WinklerPWPeter Winkler
- Brigitte ZurlBZBrigitte Zurl
- Helmuth GussHGHelmuth Guss
- Peter KindlPKPeter Kindl
- Georg StuecklschweigerGSGeorg Stuecklschweiger
- 26 January 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (4) , 643-654
Abstract
A system for dosimetric verification of intensity-modulated radiotherapy (IMRT) treatment plans using absolute calibrated radiographic films is presented. At our institution this verification procedure is performed for all IMRT treatment plans prior to patient irradiation. Therefore clinical treatment plans are transferred to a phantom and recalculated. Composite treatment plans are irradiated to a single film. Film density to absolute dose conversion is performed automatically based on a single calibration film. A software application encompassing film calibration, 2D registration of measurement and calculated distributions, image fusion, and a number of visual and quantitative evaluation utilities was developed. The main topic of this paper is a performance analysis for this quality assurance procedure, with regard to the specification of tolerance levels for quantitative evaluations. Spatial and dosimetric precision and accuracy were determined for the entire procedure, comprising all possible sources of error. The overall dosimetric and spatial measurement uncertainties obtained thereby were 1.9% and 0.8 mm respectively. Based on these results, we specified 5% dose difference and 3 mm distance-to-agreement as our tolerance levels for patient-specific quality assurance for IMRT treatments.Keywords
This publication has 21 references indexed in Scilit:
- Verification of IMRT: Techniques and ProblemsStrahlentherapie und Onkologie, 2004
- Dosimetry of therapeutic photon beams using an extended dose range filmMedical Physics, 2002
- Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimatorMedical Physics, 2002
- Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluationMedical Physics, 2002
- Variation of sensitometric curves of radiographic films in high energy photon beamsMedical Physics, 2001
- The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetryMedical Physics, 1999
- A software tool for the quantitative evaluation of 3D dose calculation algorithmsMedical Physics, 1998
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- Hierarchical chamfer matching: a parametric edge matching algorithmIEEE Transactions on Pattern Analysis and Machine Intelligence, 1988
- Dosimetric evaluation of a pencil-beam algorithm for electrons employing a two-dimensional heterogeneity correctionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1984