Evaluation of a point‐of‐care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows
- 27 March 2017
- journal article
- research article
- Published by Wiley in Australian Veterinary Journal
- Vol. 95 (4) , 123-128
Abstract
Objectives: To evaluate and validate a hand‐held electrochemical meter (Precision Xtra®) as a screening test for subclinical ketosis and hypoglycaemia in lactating dairy cattle.Design: Method comparison study using a convenience sample.Procedure: Blood samples were collected into plain tubes from the coccygeal vessels of 181 Holstein cows at 2–4 weeks of lactation during summer in Iran. Blood β‐hydroxybutyrate concentration (BHB) and glucose concentration were immediately measured by the electrochemical meter after applying 20 μL of blood to the reagent strip. Passing–Bablok regression and Bland‐Altman plots were used to determine the accuracy of the meter against laboratory reference methods (BHB dehydrogenase and glucose oxidase).Results: Serum BHB ranged from 0.1 to 7.3 mmol/L and serum glucose ranged from 0.9 to 5.1 mmol/L. Passing–Bablok regression analysis indicated that the electrochemical meter and reference methods were linearly related for BHB and glucose, with a slope estimate that was not significantly different from 1.00. Clinically minor, but statistically significant, differences were present for the intercept value for Passing–Bablok regression analysis for BHB and glucose, and bias estimates in the Bland‐Altman plots for BHB and glucose.Conclusion: The electrochemical meter provided a clinically useful method to detect subclinical ketosis and hypoglycaemia in lactating dairy cows. Compared with other method validation studies using the meter, we attributed the improved performance of the electrochemical meter to application of a fixed volume of blood (20 μL) to the reagent strip, use of the meter in hot ambient conditions and use of glucose oxidase as the reference method for glucose analysis.Keywords
This publication has 31 references indexed in Scilit:
- Bovine subclinical ketosis in dairy herds in Nishaboor, IranComparative Clinical Pathology, 2011
- Use of a hand-held meter for detecting subclinical ketosis in dairy cowsResearch in Veterinary Science, 2010
- Evaluation of a Portable Meter to Measure Ketonemia and Comparison with Ketonuria for the Diagnosis of Canine Diabetic KetoacidosisJournal of Veterinary Internal Medicine, 2009
- The Technology Behind Glucose Meters: Test StripsDiabetes Technology & Therapeutics, 2008
- Bedside Monitoring of Blood β -Hydroxybutyrate Levels in the Management of Diabetic Ketoacidosis in ChildrenDiabetes Technology & Therapeutics, 2006
- Changes in blood β-hydroxybutyrate and glucose concentrations during dry and lactation periods in Iranian Holstein cowsComparative Clinical Pathology, 2006
- Monitoring and testing dairy herds for metabolic diseaseVeterinary Clinics of North America: Food Animal Practice, 2004
- Bedside ketone determination in diabetic children with hyperglycemia and ketosis in the acute care settingPediatric Diabetes, 2004
- Evaluation of the Accusport®Lactate AnalyserInternational Journal of Sports Medicine, 1998
- Concentrations of Blood and Milk Ketone Bodies, Blood Isopropanol and Plasma Glucose in Dairy Cows in Relation to the Degree of Hyperketonaemia and Clinical Signs*Zentralblatt für Veterinärmedizin Reihe A, 1984