Technical Aspects of the WaveLight FS200 Femtosecond Laser
Authors:
- Michael MrochenMMMichael Mrochen
- Christian WüllnerCWChristian Wüllner
- Johannes KrauseJKJohannes Krause
- Mario KlafkeMKMario Klafke
- Christof DonitzkyCDChristof Donitzky
- Theo SeilerTSTheo Seiler
- 1 October 2010
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 26 (10) , S833-40
Abstract
Purpose: To demonstrate the feasibility and technical features of the WaveLight FS200 femtosecond laser (WaveLight GmbH) in refractive and corneal laser surgery. Methods: The reproducibility of femtosecond laser–created flap dimensions and channels used for corneal ring segment implantation were measured by means of an optical coherence tomography system in porcine corneas. The increase in the intraocular pressure (IOP) during applanation with the corneal interface was measured by means of an IOP sensor. The possibility of creating a channel for gas diffusion within the stroma was investigated to minimize opaque bubble layer formation. Results: The technical features, such as optical design, pulse energy, and scanning algorithms at a high repetition rate, allow reproducibility for tissue cutting with a standard deviation of <10 μm in depth and 0.1 mm laterally. Intraocular pressure increases up to 150 mmHg during applanation. The use of a gas channel minimizes the creation of an opaque bubble layer during the femtosecond laser procedure. Conclusions: The WaveLight FS200 femtosecond laser provides a high reproducibility for cutting three-dimensional tissue structures. The IOP increase is comparable to other microkeratomes and femtosecond lasers. The minimized opaque bubble layers allow the surgeon to perform an excimer laser treatment immediately after FS200 flap creation.Keywords
This publication has 20 references indexed in Scilit:
- In Vivo Real-Time Intraocular Pressure Variations during LASIK Flap CreationInvestigative Ophthalmology & Visual Science, 2010
- Second femtosecond laser pass for incomplete laser in situ keratomileusis flaps caused by suction lossJournal of Cataract & Refractive Surgery, 2009
- Incidence, possible risk factors, and potential effects of an opaque bubble layer created by a femtosecond laserJournal of Cataract & Refractive Surgery, 2008
- Characterization of submicrojoule femtosecond laser corneal tissue dissectionJournal of Cataract & Refractive Surgery, 2008
- Advances in refractive surgery: microkeratome and femtosecond laser flap creation in relation to safety, efficacy, predictability, and biomechanical stabilityCurrent Opinion in Ophthalmology, 2006
- One thousand consecutive IntraLase laser in situ keratomileusis flapsJournal of Cataract & Refractive Surgery, 2006
- Femtosecond Laser Preparation of Donor Tissue From the Endothelial SideCornea, 2006
- A Practical Femtosecond Laser Procedure for DLEK Endothelial TransplantationCornea, 2005
- Femtosecond Laser Posterior Lamellar KeratoplastyCornea, 2005
- Reproducibility of corneal flap thickness in laser in situ keratomileusis using the Hansatome microkeratomeJournal of Cataract & Refractive Surgery, 2000