Determination of chromium( iii ) by differential pulse stripping voltammetry at a chitosan–gold nanocomposite modified screen printed electrode

Authors:

Abstract
Detection of trivalent chromium with screen printed electrodes modified by chitosan–gold nanocomposites using differential pulse voltammetry (DPSV), showing a linear range from 1 to 100 μM and a detection limit of 0.4 μM (based on 3 σ ).

This publication has 33 references indexed in Scilit: