Abstract
This paper presents a novel method for monitoring drug cytotoxicity using a hybrid microfluidic CMOS platform. This platform consists of an array of 8 × 8 capacitive sensors integrated with a readout circuit on the same chip. In this paper, we present a layer-by-layer (LBL) polyelectrolyte deposition technique to coat the surface of microelectrodes realized in the top most metal layer in 0.35-μm CMOS process. This process successfully enhances the biocompatibility of sensing microelectrodes and consequently increases the cell viability over a three-day period. Herein, we demonstrate and discuss the advantage of the proposed platform for drug cytotoxicity as well as cellular growth monitoring. This CMOS sensing platform possesses a wide output dynamic range and allows tracking cell growth at initial cell concentrations ranging from 10 to 200 k Cells/ml. We also use a standard Alamarblue cell-based assay and Geneticin selective antibiotic (G418) as control and cytotoxic drugs introduced to non-resistant H1299 and resistant Hek293 cell lines, respectively. Furthermore, a low complexity microfluidic packaging technique is presented to create and bond micro-wells on CMOS chip for rapid test and characterization. With the potential to perform label-free cellular analysis, the proposed platform opens an avenue to transition from traditional to smart cellular analysis techniques suitable for a variety of biological applications, in particular high throughput cell-based drug testing.Keywords
Funding Information
- Canada Research Chair on Smart Medical Devices
- Natural Sciences and Engineering Research Council of Canada
- Carbon Management Canada
This publication has 34 references indexed in Scilit:
- A CMOS magnetic microbead-based capacitive biosensor array with on-chip electromagnetic manipulationBiosensors and Bioelectronics, 2013
- Polyelectrolyte Multilayer Assemblies on Materials Surfaces: From Cell Adhesion to Tissue EngineeringChemistry of Materials, 2011
- Cell Surface Engineering with Polyelectrolyte Multilayer Thin FilmsJournal of the American Chemical Society, 2011
- Electrical cell-substrate impedance sensing as a non-invasive tool for cancer cell studyThe Analyst, 2010
- Impedance studies of bio-behavior and chemosensitivity of cancer cells by micro-electrode arraysBiosensors and Bioelectronics, 2008
- Chip–NMR biosensor for detection and molecular analysis of cellsNature Medicine, 2008
- Tracking cancer cell proliferation on a CMOS capacitance sensor chipBiosensors and Bioelectronics, 2008
- Formation and Dielectric Properties of Polyelectrolyte Multilayers Studied by a Silicon-on-Insulator Based Thin Film ResistorLangmuir, 2007
- Functional cellular assays with multiparametric silicon sensor chipsLab on a Chip, 2003
- Dielectric Properties of Polyelectrolyte MultilayersLangmuir, 2001