Abstract
It has been shown that the incidence of venous air embolism and venous carbon dioxide (CO2) embolism is high during endoscopic retrograde cholangiopancreatography (ERCP). We examined insufflating gas flow and maximum pressure produced by three types of commonly used endoscopes because we could not readily locate technical data for endoscope gas flow and maximum emitted pressure in the manufacturer's manuals. We tested the Olympus GIF-Q180 used for esophagogastroduodenoscopy, the CF-Q180 used for colonoscopy, and the TJF-Q180 used for ERCP (Olympus America Inc., Center Valley, Pennsylvania). Under three different clinical gas insufflation scenarios, we measured in vitro maximum gas pressure transduced from a closed space created at the endoscope tip in a worst-case scenario analysis. We showed that it is readily possible to generate a pressure (>5-30 times normal central venous pressure) in the air space at the tip of all three endoscopes when insufflation is activated and the gas egress is limited. These findings shed additional light on in vivo occurrences of gas embolism during gastrointestinal endoscopy. We postulate that in addition to using exclusively CO2 as the insufflating gas, the risk of gas embolism can be further diminished by regulating insufflating gas pressure at the tip of endoscopes.