Long‐term outcomes with abandoning versus extracting sterile leads: A 10‐year population‐based study
- 15 March 2024
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 47 (5) , 626-634
Abstract
Background: Long‐term outcomes of sterile lead management strategies of lead abandonment (LA) or transvenous lead extraction (TLE) remain unclear. Methods: We performed a retrospective study of a population residing in southeastern Minnesota with follow‐up at the Mayo Clinic and its health systems. Patients who underwent LA or TLE of sterile leads from January 1, 2000, to January 1, 2011, and had follow‐up for at least 10 years or until their death were included. Results: A total of 172 patients were included in the study with 153 patients who underwent LA and 19 who underwent TLE for sterile leads. Indications for subsequent lead extraction arose in 9.1% ( n = 14) of patients with initial LA and 5.3% ( n = 1) in patients with initial TLE, after an average of 7 years. Moreover, 28.6% of patients in the LA cohort who required subsequent extraction did not proceed with the extraction, and among those who proceeded, 60% had clinical success and 40% had a clinical failure. Subsequent device upgrades or revisions were performed in 18.3% of patients in the LA group and 31.6% in the TLE group, with no significant differences in procedural challenges (5.2% vs. 5.3%). There was no difference in 10‐year survival probability among the LA group and the TLE group ( p = .64). Conclusion: An initial lead abandonment strategy was associated with more complicated subsequent extraction procedures compared to patients with an initial transvenous lead extraction strategy. However, there was no difference in 10‐year survival probability between both lead management approaches.Keywords
This publication has 23 references indexed in Scilit:
- Long-term survival following transvenous lead extraction: Importance of indication and comorbiditiesHeart Rhythm, 2021
- Pacemaker lead-associated tricuspid regurgitation in patients with or without pre-existing right ventricular dilatationClinical Research in Cardiology, 2021
- Trends of Cardiovascular Implantable Electronic Device Infection in 3 DecadesJACC: Clinical Electrophysiology, 2019
- Procedural outcomes and long-term survival associated with lead extraction in patients with abandoned leadsHeart Rhythm, 2018
- 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extractionHeart Rhythm, 2017
- Cardiac Implantable Electronic Device InfectionsJACC: Clinical Electrophysiology, 2017
- Tricuspid Regurgitation and Mortality in Patients With Transvenous Permanent Pacemaker LeadsThe American Journal of Cardiology, 2016
- To Extract or Retain a Sterile, Nonfunctional LeadCardiac Electrophysiology Clinics, 2015
- Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009JACC, 2012
- Incidence and Risk Factors of Upper Extremity Deep Vein Lesions After Permanent Transvenous Pacemaker Implant: A 6‐Month Follow‐Up Prospective StudyPacing and Clinical Electrophysiology, 2002