Improved fibre optic respiratory monitoring using a figure-of-eight coil
Authors:
- F-X MaletrasFMF-X Maletras
- J MasonJMJ Mason
- 31 May 2005
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 26 (5) , 585-590
Abstract
The macro-bending loss effect in optical fibres has been used to redevelop a sensor for the measurement of thoracic and abdominal circumferences in non-invasive respiratory monitoring. The new sensor uses a novel figure-of-eight loop configuration, which results in increased linearity of response, less mechanical resistance and hysteresis, as well as other benefits. The performance of the new sensor as applied to respiratory monitoring is examined, and indicates a higher resolution and sensitivity than the old. This enhanced performance enables measurement of respiratory and cardiac function using the same transducing fibre.Keywords
This publication has 6 references indexed in Scilit:
- A fibre optic Bragg grating strain sensor for monitoring ventilatory movementsMeasurement Science and Technology, 2001
- A new sensor for monitoring chest wall motion during high-frequency oscillatory ventilationMedical Engineering & Physics, 1999
- A simple fiber-optic sensor for use over a large displacement rangeOptics and Lasers in Engineering, 1997
- A theoretical study of the robustness of the isovolume calibration method for a two-compartment model of breathing, based on an analysis of the connected cylinders modelPhysics in Medicine & Biology, 1997
- Design and characterization of a fiber optic respiratory plethysmograph (FORP)1996
- Measurement of the separate volume changes of rib cage and abdomen during breathingJournal of Applied Physiology, 1967