Abstract
An experiment was carried out to demonstrate the detection of a hydrogen emission line, HI656.2nm (Hα) , in a plasma induced by a Q -switched Nd-YAG (YAG, yttrium aluminium garnet) laser in a low pressure gas on various types of samples, such as zinc, a glass slide, and a zircalloy tube. Contribution by surface water could be suppressed by a laser cleaning treatment and the resulting calibration curve obtained for zircalloy tube samples doped with various concentrations of hydrogen (0 , 200 , 540 , and 960 ) suggest potential applications to the quantitative analysis of hydrogen. A study of the dynamic process represented by the time profiles of the hydrogen emission, in comparison with those for zinc atomic emission, revealed a specific feature that is related to the small mass of hydrogen. This specific feature can be explained by the shock wave excitation mechanism in terms of new hypothetical process, namely, a mismatch between the movement of ablated hydrogen atoms and the formation of the shock wave.

This publication has 9 references indexed in Scilit: