A Novel Compact Pneumatic Turbine-Driven Transducer for MR Elastography
Authors:
- Lorenz KissLKLorenz Kiss
- Jean-Lynce GnanagoJGJean-Lynce Gnanago
- Stefan WamplSWStefan Wampl
- Marcos WolfMWMarcos Wolf
- Christopher BreitenfelderCBChristopher Breitenfelder
- Andreas HodulAHAndreas Hodul
- Quang NguyenQNQuang Nguyen
- Markus KönigshoferMKMarkus Königshofer
- Roberta Frass-KrieglRFRoberta Frass-Kriegl
- Martin MeyerspeerMMMartin Meyerspeer
- Albrecht Ingo SchmidASAlbrecht Ingo Schmid
- 8 June 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 75 (00189456) , 4505409
Abstract
Magnetic Resonance Elastography (MRE) can be used to estimate biomechanical parameters, e.g., shear stiffness of human tissue. As such, it is an imaging-based extension of palpation. To perform MRE, mechanical waves are induced into the tissue by transducers or actuators. The resulting periodic microscopic tissue displacements are recorded using phase-contrast magnetic resonance imaging (MRI). Good transducers need sufficient vibration amplitudes and frequency stability with little contributions of higher order harmonics. For clinical use, it needs to be compact for comfort as well as interference with RF coil placement. Safe operation in strong electromagnetic fields is a must. Setup procedures should be simple and quick. Here, a new compact pneumatically driven gravitational transducer (CPT) for MR elastography is presented. It consists of rotating unbalances stacked on either side of a turbine. Additional insertable masses allow adjustment of the vibration amplitude at a set frequency. Additive manufacturing and milling were combined to produce the transducer. The flat design of the CPT allows for placement between body and receiver coil with minimal sensitivity loss in most relevant targets. PID control is used to precisely control vibration frequency, based on optical detection of rotation cycles. It adjusts a valve in the air supply of the turbine. It is constructed using non-magnetic, non-conducting materials, which makes it inherently MRI-safe. The electronics, fittings and valves are placed in a control unit (CU) outside the magnet room. A hose and two optical fibres connect the transducer and the CU. The performance of the transducer was thoroughly bench tested and phantom experiments in the MRI scanner demonstrated the feasibility and robustness of the system. The frequency control was excellent for a pneumatic system (FWHM 0.04-0.20 Hz) across a wide range (50-130 Hz), while amplitudes of higher harmonics are very low (1.9-8.1 % 1st, 0.2-4.6 % 2nd order). Imaging data confirm that the CPT generated displacements which could be recorded with MR elastography pulse sequences. In conclusion, we present a new MRE pneumatically driven gravitational transducer, Its advantages are the combination of the following features: It requires no electrical components in the vicinity of the MR scanner. It performs equally well as electromechanically driven systems while it outperforms pneumatic designs in terms of frequency stability. Its compact design allows for flexibility in placement with a variety of RF coils and target tissues in the body. It consists of low-cost components and is easily used. We conclude that the combination of the CPT’s features constitute a substantial improvement in MRE transducer design.Keywords
Funding Information
- Austrian Science Fund (DOI: 10.55776/P35607)
This publication has 47 references indexed in Scilit:
- Biomechanical Assessment of Liver Integrity: Prospective Evaluation of Mechanical Versus Acoustic MR ElastographyJournal of Magnetic Resonance Imaging, 2024
- Comparison of inversion methods inMRelastography: An open‐access pipeline for processing multifrequency shear‐wave data and demonstration in a phantom, human kidneys, and brainMagnetic Resonance in Medicine, 2022
- Design, Construction, and Implementation of a Magnetic Resonance Elastography Actuator for Research PurposesCurrent Protocols, 2022
- The Accuracy of iPhone Applications to Monitor Environmental Noise LevelsThe Laryngoscope, 2020
- Ristretto MRE: A generalized multi‐shot GRE‐MRE sequenceNMR in Biomedicine, 2019
- Cardiovascular magnetic resonance elastography: A reviewNMR in Biomedicine, 2017
- Magnetic resonance elastography: A reviewClinical Anatomy, 2010
- Comprehensive quantification of signal‐to‐noise ratio and g‐factor for image‐based and k‐space‐based parallel imaging reconstructionsMagnetic Resonance in Medicine, 2008
- Electromagnetic actuator for generating variably oriented shear waves in MR elastographyMagnetic Resonance in Medicine, 2003
- Complex-valued stiffness reconstruction for magnetic resonance elastography by algebraic inversion of the differential equationMagnetic Resonance in Medicine, 2001