Addressing recent challenges in isotope ratio mass spectrometry: Development of a method applicable to 1‐androstene‐steroids, 6α‐hydroxy‐androstenedione, and androstatrienedione
Open Access
- 31 August 2022
- journal article
- research article
- Published by Wiley in Drug Testing and Analysis
- Vol. 14 (11-12) , 1891-1903
Abstract
In 2020, the confirmation of the non‐endogenous origin of several pseudo‐endogenous steroids by means of isotope ratio mass spectrometry (IRMS) was recommended by the World Anti‐Doping Agency (WADA), in addition to previously established target analytes for IRMS in sports drug testing. To date, however, IRMS‐based methods validated in accordance with current WADA regulations have not been available. Therefore, the aim of this research project was the development and validation of a method to determine the carbon isotope ratios (CIR) of all newly considered pseudo‐endogenous steroids, encompassing the anabolic androgenic steroids comprising a 1‐ene‐core structure (5α‐androst‐1‐ene‐3β,17β‐diol, 5α‐androst‐1‐ene‐3,17‐dione [1AD], 17β‐hydroxy‐5α‐androst‐1‐en‐3‐one, 3α‐hydroxy‐5α‐androst‐1‐ene‐17‐one [1AND], and 3β‐hydroxy‐5α‐androst‐1‐ene‐17‐one [1EpiAND]), as well as steroids referred to as hormone and metabolic modulators (androsta‐1,4,6‐triene‐3,17‐dione [TRD] and its main metabolite 17β‐hydroxy‐androsta‐1,4,6‐triene‐3‐one) and 6α‐ and 6β‐hydroxy‐androst‐4‐ene‐3,17‐dione. With peak purity of target analytes being critical for IRMS analyses, a twofold high‐performance liquid chromatography (HPLC)‐based sample purification was employed, with all analytes being acetylated between the first and second HPLC fractionation. Using established gas chromatography/combustion/IRMS instrumentation, limits of quantification were estimated at 10 ng/ml for a 20 ml urine aliquot for all analytes, except for 1AND (20 ng/ml), and combined measurement uncertainties were estimated between 0.4‰ and 0.9‰. For proof‐of‐concept, samples collected after the single oral administration of a nutritional supplement containing 1AD and 1EpiAND were analyzed as well as existing excretion study urine samples obtained after the administration of 4‐androstenedione and TRD. Based on the obtained results, the developed method was considered to be fit‐for‐purpose.Keywords
Funding Information
- Bundesministerium des Innern, für Bau und Heimat
- World Anti-Doping Agency (20A07MT)
This publication has 17 references indexed in Scilit:
- Studies on the minor metabolite 6a‐hydroxy‐androstenedione for doping control purposes and its contribution to the steroid profileDrug Testing and Analysis, 2014
- Investigations on hydrogen isotope ratios of endogenous urinary steroids: reference‐population‐based thresholds and proof‐of‐conceptDrug Testing and Analysis, 2012
- Seized designer supplement named “1-Androsterone”: Identification as 3β-hydroxy-5α-androst-1-en-17-one and its urinary eliminationSteroids, 2011
- Determination of 13C/12C ratios of urinary excreted boldenone and its main metabolite 5β‐androst‐1‐en‐17β‐ol‐3‐oneDrug Testing and Analysis, 2010
- Steroid isotopic standards for gas chromatography-combustion isotope ratio mass spectrometry (GCC-IRMS)Steroids, 2009
- Metabolism of androsta‐1,4,6‐triene‐3,17‐dione and detection by gas chromatography/mass spectrometry in doping controlRapid Communications in Mass Spectrometry, 2008
- Factors influencing the steroid profile in doping control analysisJournal of Mass Spectrometry, 2008
- Determination of 13 C/ 12 C ratios of endogenous urinary steroids: method validation, reference population and application to doping control purposesRapid Communications in Mass Spectrometry, 2008
- Isotopic Fractionation of Endogenous Anabolic Androgenic Steroids and Its Relationship to Doping Control in SportsJournal of Chromatographic Science, 2005