Abstract
Chemometric algorithms are commonly applied in spectroscopy for deriving concentrations of chemicals from spectra. In order to ensure correct concentration results, the background spectrum must be stable and all absorbers present in unknown samples must have been contained in the calibration set. Recently, standard principal component regression (PCR) has been improved in two ways, correcting either for spectral drifts or for uncalibrated spectral features. In this study a combination of both correction approaches named secured pseudo PCR (spPCR) is presented which provides substantially improved concentration results even if the spectra are simultaneously affected by drifts and uncalibrated features. Additionally, spPCR extracts estimates of the spectrum of the uncalibrated absorbers and the drift spectrum. Hence systematic disturbances can be analyzed and a facile tool for allocating their origin is provided. Copyright © 2004 John Wiley & Sons, Ltd.