PET-based dose delivery verification in proton therapy: a GATE based simulation study of five PET system designs in clinical conditions
Authors:
- Charlotte RobertCharlotte Robert
- David SarrutDSDavid Sarrut
- Simon StuteSSSimon Stute
- Irène BuvatIBIrène Buvat
- 12 September 2013
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (19) , 6867-6885
Abstract
PET is a promising technique for in vivo treatment verification in hadrontherapy. Three main PET geometries dedicated to in-beam treatment monitoring have been proposed in the literature: the dual-head PET geometry, the OpenPET geometry and the slanted-closed ring geometry. The aim of this work is to characterize the performance of two of these dedicated PET detectors in realistic clinical conditions. Several configurations of the dual-head PET and OpenPET systems were simulated using GATE v6.2. For the dual-head configuration, two aperture angles (15° and 45°) were studied. For the OpenPET system, two gaps between rings were investigated (110 and 160 mm). A full-ring PET system was also simulated as a reference. After preliminary evaluation of the sensitivity and spatial resolution using a Derenzo phantom, a real small-field head and neck treatment plan was simulated, with and without introducing patient displacements. No wash-out was taken into account. 3D maps of the annihilation photon locations were deduced from the PET data acquired right after the treatment session (5 min acquisition) using a dedicated OS-EM reconstruction algorithm. Detection sensitivity at the center of the field-of-view (FOV) varied from 5.2% (45° dual-head system) to 7.0% (full-ring PET). The dual-head systems had a more uniform efficiency within the FOV than the OpenPET systems. The spatial resolution strongly depended on the location within the FOV for the ϕ = 45° dual-head system and for the two OpenPET systems. All investigated architectures identified the magnitude of mispositioning introduced in the simulations within a 1.5 mm accuracy. The variability on the estimated mispositionings was less than 2 mm for all PET systems.Keywords
This publication has 43 references indexed in Scilit:
- In-beam PET measurements of biological half-lives of12C irradiation induced β+-activityActa Oncologica, 2008
- Direct time-of-flight for quantitative, real-time in-beam PET: a concept and feasibility studyPhysics in Medicine & Biology, 2007
- The FLUKA code: description and benchmarkingAIP Conference Proceedings, 2007
- On the detector arrangement for in-beam PET for hadron therapy monitoringPhysics in Medicine & Biology, 2006
- Charged hadron tumour therapy monitoring by means of PETNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Investigation of scattered radiation in 3D whole-body positron emission tomography using Monte Carlo simulationsPhysics in Medicine & Biology, 1999
- Positron emission tomography for quality assurance of cancer therapy with light ion beamsNuclear Physics A, 1999
- Object shape dependent scatter simulations for PETIEEE Transactions on Nuclear Science, 1991
- High Resolution Computed Tomography of Positron EmittersIEEE Transactions on Nuclear Science, 1977