Transiently Increased Variation Between a Point-of-Care and Laboratory INR Method After a Long Period of Correlation
Open Access
- 1 October 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 140 (4) , 475-486
Abstract
To perform long-term comparison between laboratory Stago and Point-of-Care (POC) i-STAT methods for determining the international normalized ratio (INR). This was a multicenter method comparison of patient INR results and factors related to performance variance. For 5 years, the assays demonstrated close patient correlation within and above the 3.5 INR therapeutic range cutoff (bias, 0.23 INR units). Patient results above 3.5 INR were bimodal, with 60% demonstrating an i-STAT INR bias of less than 0.5. Several patient conditions were associated with the presence of a higher i-STAT bias. In year 6, a broader range i-STAT bias developed, increasing to 0.73 INR units. The increased bias persisted for 3 years, then returned to initial levels following i-STAT adjustments. The substantial increase in i-STAT bias after a long period of stability was partly corrected by renewed correlation to the international reference preparation. Additional assay drift is discussed in relation to thromboplastin reagents and other testing variables. This study emphasizes the need for continual laboratory correlation with POC devices and caution in using published comparisons.Keywords
This publication has 38 references indexed in Scilit:
- Comparison of International Normalized Ratios provided by two point-of-care devices and laboratory-based venipuncture in a pharmacist-managed anticoagulation clinicAmerican Journal of Health-System Pharmacy, 2010
- Comparison of Plasma With Whole Blood Prothrombin Time and Fibrinogen on the Same InstrumentAmerican Journal of Clinical Pathology, 2010
- Clinical decision-making as the basis for assessing agreement between measures of the International Normalized RatioJournal of Thrombosis and Haemostasis, 2009
- Pharmacology and Management of the Vitamin K AntagonistsChest, 2008
- The international normalized ratio to prioritize patients for liver transplantation: problems and possible solutionsJournal of Thrombosis and Haemostasis, 2008
- Influence of haematocrit on international normalised ratio (INR) differences between a whole blood point-of-care coagulation monitor and reference prothrombin time in plasmaThrombosis and Haemostasis, 2008
- European Concerted Action on Anticoagulation. A multicentre calibration study of WHO international reference preparations for thromboplastin, rabbit (RBT/90) and human (rTF/95)Journal of Clinical Pathology, 2005
- Importance of Device Evaluation for Point-of-Care Prothrombin Time International Normalized Ratio Testing ProgramsMayo Clinic Proceedings, 2005
- Point-of-care testing of the international normalized ratio in patients with antiphospholipid antibodiesThrombosis and Haemostasis, 2005
- Thromboplastin-sensitivity, precision and other characteristicsClinical and Laboratory Haematology, 1988