Method for in-shoe shear stress measurement
Authors:
- R. HoseinRHR. Hosein
- R.B. WilliamsRWR.B. Williams
- 1 May 1992
- journal article
- Published by IOP Publishing in Journal of Biomedical Engineering
- Vol. 14 (3) , 181-186
Abstract
A methodology is described for use of a shear transducer, based on a magneto-resistive principle, to measure stresses under the plantar surface of the foot in-shoe during walking. Particular attention is paid to a projected application for study of diabetic plantar ulceration and its management by footwear. The transducer has a disc construction, approximately 4 mm thick by 16 mm diameter, and measures two orthogonal axes of shear simultaneously; this disc is mounted into an inlay that can be inserted into any stock orthopaedic shoe of the type commonly prescribed for diabetic foot problems. The transducer is located in the metatarsal head region of the inlay; exact placement of the transducer is determined by reference to the direct pressure distribution, the common method of palpation shown to be imprecise. Pilot trials on normal subjects are presented to evaluate the method.Keywords
This publication has 13 references indexed in Scilit:
- Triaxial force transducer for investigating stresses at the stump/socket interfaceMedical & Biological Engineering & Computing, 1992
- Study of the relative timing of shear forces on the sole of the forefoot during walkingJournal of Biomedical Engineering, 1991
- Foot pressure patterns during gaitJournal of Biomedical Engineering, 1985
- Plantar pressure measurements and the prevention of ulceration in the diabetic footThe Journal of Bone and Joint Surgery. British volume, 1985
- Method of healing diabetic forefoot ulcers.BMJ, 1983
- Forces under the footJournal of Biomedical Engineering, 1983
- Dynamic Foot Pressure and Other Studies as Diagnostic and Management Aids in Diabetic NeuropathyDiabetes Care, 1983
- Vertical forces acting on the feet of diabetic patients with neuropathic ulcerationBritish Journal of Surgery, 1981
- Method of measuring 'shearing' forces on the sole of the footClinical Physics and Physiological Measurement, 1980
- TREATMENT OF PERFORATING ULCER OF THE FOOT.The Lancet, 1884