Burn depth assessment guides treatment selection, yet clinical evaluation remains largely visual and subjective. This simulation study investigates whether the electromagnetic signature of a microstrip-patch probe can distinguish burn stages, and which frequency band is best suited to the task. A five-state layered dielectric phantom (Healthy, Superficial, Superficial-Partial, Deep-Partial, and Full-Thickness) was interrogated in a full-wave solver by three patch probes at 2.45, 5.8, and 24 GHz, all at a common 2 mm probe-to-skin gap. At 5.8 GHz the three shallow states are nearly indistinguishable (|ΔS11| ≤ 1.7 dB relative to healthy), and at 2.45 GHz they are only modestly separated. At 24 GHz, however, the Superficial-Partial state produces a distinct resonance (S11 = -45.2 dB; ΔS11 = -14.8 dB). The 24 GHz probe is therefore the only one sensitive to the shallow stages, and it was selected for raster scanning. A boustrophedon scan of a mixed-pathology concentric-ring phantom, simulated at 441 positions on a 5 mm grid, reproduces the ring layout, recovers the four ring boundaries to within the 5 mm sampling pitch, and confirms the phantom's radial symmetry to within ~0.05 dB at matched radii. All results are simulation-based; physical phantom measurement is the next step.