Optimization of a monolithic detector block design for a prototype human brain PET scanner
- 1 October 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10823654,p. 4927-4930
Abstract
We are presently developing a novel PET scanner for human brain functional imaging based on monolithic scintillator crystals read by APD matrices, capable of being inserted into an MRI system. In this work we report on the detailed study that has been made of the design of the detector blocks, aiming at defining the most suitable geometrical and readout configuration for optimizing the overall performance of the entire scanner. Both parallel and trapezoidal geometries have been simulated, featuring two layers of active scintillator material with different or similar thickness and APD readout on the front or back side. Results of this study indicate that a trapezoidal geometry with equal thickness of both layers is the best solution for improving the expected scanner performance.Keywords
This publication has 11 references indexed in Scilit:
- A dedicated ASIC front-end readout for the monolithic detector blocks of the BrainPET prototype2008
- GAMOS: A Geant4-based easy and flexible framework for nuclear medicine applications2008
- Simultaneous PET-MRI: a new approach for functional and morphological imagingNature Medicine, 2008
- Towards a continuous crystal APD-based PET detector designNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Evaluation of monolithic detector blocks for high- sensitivity PET imaging of the human brain2007
- Positron emission tomographyPhysics in Medicine & Biology, 2006
- Neural network-based position estimators for PET detectors using monolithic LSO blocksIEEE Transactions on Nuclear Science, 2004
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Magnetic resonance imaging-guided attenuation and scatter corrections in three-dimensional brain positron emission tomographyMedical Physics, 2003
- Use of a magnetic field to increase the spatial resolution of positron emission tomographyMedical Physics, 1994