Pressure‐assisted tryptic digestion using a syringe

Authors:

Abstract
A simple and effective digestion method was developed using a syringe. A 3 mL syringe was used to apply a pressure of 6 atm to expedite tryptic digestion. Application of a pressure of 6 atm during digestion resulted in better digestion efficiency than digestion under atmospheric pressure. The protein peaks in the matrix‐assisted laser desorption/ionization mass spectra of three model proteins (cytochrome c, horse heart myoglobin, and bovine serum albumin (BSA)) completely disappeared within 30 min at 37°C under a pressure of 6 atm, with greater numbers of peptides observed in 30 min pressure‐assisted digestion than in overnight atmospheric pressure digestion. This is mostly due to the miscleaved peptides. Similar sequence coverages were obtained for 30 min pressure‐assisted digestion and overnight atmospheric pressure digestion of the three model proteins (92% vs. 88% for cytochrome c , 100% vs. 97% for horse heart myoglobin, and 53% vs. 53% for BSA). Copyright © 2010 John Wiley & Sons, Ltd.

This publication has 33 references indexed in Scilit: