Risk Factors for False Positive Transcranial Motor Evoked Potential Monitoring Alerts During Surgical Treatment of Cervical Myelopathy
Authors:
- David H. KimDKDavid H. Kim
- Jason ZaremskiJZJason Zaremski
- Brian KwonBKBrian Kwon
- Louis JenisLJLouis Jenis
- Eric WoodardEWEric Woodard
- Robert BodeRBRobert Bode
- Robert J. BancoRBRobert J. Banco
- 1 December 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 32 (26) , 3041-3046
Abstract
Retrospective consecutive series review. To examine performance of transcranial motor-evoked potential (TcMEP) monitoring in patients undergoing surgery for cervical myelopathy and potential risk factors for false positive alerts. Although use of TcMEP monitoring has been increasing and has been specifically recommended in patients with cervical myelopathy, rates and risk factors for false positive alerts have not been established. Intraoperative neuromonitoring data for 52 consecutive patients undergoing surgery for cervical myelopathy were reviewed. All major TcMEP alerts were identified. Comprehensive demographic and clinical data, preoperative imaging studies, operative, and anesthesia records were reviewed. Six of 52 patients (12%) experienced a major TcMEP alert consisting of sustained >80% loss of amplitude. There were no somatosensory-evoked potential (SSEP)-related alerts. In 2 cases, an intraoperative wake-up test was negative and in 3 cases, surgery was completed without a wake-up test and without recovery of TcMEP signals. No new postoperative neurologic deficits were observed in these patients. One patient with new postoperative weakness was correctly predicted by loss of TcMEP signals. No new deficit was observed in the remaining 46 patients. Statistical analysis revealed significantly higher body mass index (28.8 vs. 35.0; P = 0.032) and length of surgery (191 vs. 283 minutes; P = 0.019) in patients with false positive alerts. In this series of cervical myelopathy patients, sensitivity and specificity of TcMEP for detection of clinically significant intraoperative cord injury were 100% and 90%, respectively. Sensitivity and specificity of SSEP were 0% and 100%, respectively. The positive predictive value of a TcMEP alert was 17%. Possible risk factors for false positive TcMEP alerts include obesity and increased length of surgery. This study supports superior sensitivity of TcMEP compared with SSEP monitoring but identifies a relatively high false positive rate even in a selected high-risk cervical myelopathy population when this modality is applied in practice.Keywords
This publication has 11 references indexed in Scilit:
- Characterization of Neurophysiologic Alerts During Anterior Cervical Spine SurgerySpine, 2006
- Comparison of Transcranial Electric Motor and Somatosensory Evoked Potential Monitoring During Cervical Spine SurgeryJournal of Bone and Joint Surgery, 2004
- Neurophysiological monitoring of spinal cord function during instrumented anterior cervical fusionThe Spine Journal, 2004
- Two cases of quadriparesis following anterior cervical discectomy, with normal perioperative somatosensory evoked potentialsJournal of Neurology, Neurosurgery & Psychiatry, 2003
- Combined monitoring of motor and somatosensory evoked potentials in orthopaedic spinal surgeryClinical Neurophysiology, 2002
- Intraoperative control by somatosensory evoked potentials in the treatment of cervical myeloradiculopathy. Results in 210 casesEuropean Spine Journal, 1997
- Intraoperative Monitoring of Spinal Cord Function Using Motor Evoked Potentials via Transcutaneous Epidural Electrode During Anterior Cervical Spinal SurgeryJournal of Spinal Disorders, 1997
- Intraoperative monitoring of motor evoked potentialsNeurology, 1996
- Somatosensory evoked potential monitoring in cervical surgery: identification of pre- and intraoperative risk factors associated with neurological deteriorationJournal of Neurosurgery, 1996
- Motor Evoked Potential Monitoring during Upper Cervical Spine SurgerySpine, 1989