FIx Controller: An Instrument to Automatically Adjust Inspired Oxygen Fraction Using Feedback Control from a Pulse Oximeter

Authors:

Abstract
Objective. To develop an instrument to help prevent pulmonaryO2 toxicity, a syndrome that manifests itself in adultintensive care patients. Methods. We designed, built, and tested adevice that controls FIO2 exposure using oxygensaturation measured with a pulse oximeter (SpO2) in a negativefeedback control system. A target SpO2 is designated by theclinician and the system adjusts the FIO2 from amechanical ventilator so as to minimize the difference between the measuredSpO2 and the target. Important elements of the system includea conservative artifact rejection algorithm, a gain-scheduled sampled-dataproportional-integral-derivative (PID) controller, and a safety system toprevent inspired mixtures with undesirably low FIO2 dueto device failure. Results. The control system was tuned in a series ofanimal experiments. Acceptable clinical response of the system was obtainedusing a gain-scheduled controller algorithm whereby the gain of theproportional term of a PID controller was adjusted based on the error signaland measured minute ventilation. Also, the artifact rejection algorithm andsafety systems were successfully tested using simulation. Conclusions.Testing the effectiveness of this instrument will require comparison withmanual control of FIO2 in an appropriately designedtrial.