Preliminary Investigation of a Cd0.9Zn0.1Te Detector for Small-Field Dosimetry Applications Using Therapeutic MV Beams
Open Access
- 7 February 2025
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 15 (4) , 1693
Abstract
Stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) require precise small-field dosimetry, verified through patient-specific quality assurance (PSQA). This study evaluated the feasibility of using a single-crystal cadmium–zinc–telluride (Cd0.9Zn0.1Te, CZT) detector for PSQA in SRS and SBRT. We fabricated a CZT detector with Au electrodes and examined its fundamental characteristics, including dose linearity, dose rate dependence, energy dependence, angular dependence, source-to-surface distance (SSD) dependence, field size dependence, depth dependence, and reproducibility, under 6 and 10 MV LINAC beam irradiation and compared the results with those from a standard ionization chamber. The results revealed that the CZT detector demonstrated excellent linearity across 0–1000 cGy with minimal deviation in the low-dose region, negligible dose rate dependence, and minimal energy dependence, exhibiting a 2.2% drop at 15 MV relative to 6 MV. Its angular and SSD dependencies deviated slightly from the ionization chamber, consistent with the expected physical behaviors and correctable in clinical practice. The detector also revealed consistent performance over time with excellent reproducibility, and its depth dependence results were consistent with those of the ionization chamber. Thus, the CZT detector provides consistent performance in small-field measurements under varying conditions, satisfying the requirements for SRS and SBRT.Keywords
Funding Information
- Chonnam National University Hospital Biomedical Research Institute (BCRI24053)
This publication has 35 references indexed in Scilit:
- Comparative study of SRS end‐to‐end QA processes of a diode array device and an anthropomorphic phantom loaded with GafChromic XD filmJournal of Applied Clinical Medical Physics, 2022
- Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgeryPhysics and Imaging in Radiation Oncology, 2021
- AAPM‐RSS Medical Physics Practice Guideline 9.a. for SRS‐SBRTJournal of Applied Clinical Medical Physics, 2017
- Energy and field size dependence of a silicon diode designed for small‐field dosimetryMedical Physics, 2017
- Enhancing safety and quality through preplanning peer review for patients undergoing stereotactic body radiation therapyPractical Radiation Oncology, 2016
- 2D EPID dose calibration for pretreatment quality control of conformal and IMRT fields: A simple and fast convolution approachPhysica Medica, 2015
- In Regard to Koshy et alInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Growth of CdTexSe1−x from a Te-rich solution for applications in radiation detectionJournal of Crystal Growth, 2014
- The Canadian Association of Radiation Oncology Scope of Practice Guidelines for Lung, Liver and Spine Stereotactic Body RadiotherapyClinical Oncology, 2012
- Scanning orientation effects on Gafchromic EBT film dosimetryAustralasian Physics & Engineering Sciences in Medicine, 2006