Effectiveness and Reliability of Selected Site Pacing for Avoidance of Phrenic Nerve Stimulation in CRT Patients with Quadripolar LV Leads: The EffaceQ Study
Authors:
- 12 June 2015
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 38 (8) , 942-950
Abstract
Background: Cardiac resynchronization therapy (CRT) relies on sufficient left ventricular (LV) pacing with safety margin to phrenic nerve stimulation (PNS). Previous studies introduced LV vector reprogramming in bipolar coronary sinus leads to optimize LV pacing and avoid PNS. We investigated the efficacy and reliability of quadripolar leads in CRT.Methods: The EffaceQ study enrolled 344 patients with de novo CRT implantation with a quadripolar LV lead in an observational, prospective multicenter study. The study was powered to demonstrate that in at least 90% of patients with an implanted quadripolar LV lead, a viable LV pacing configuration (LVPC) is available (primary end point: LV pacing threshold ≤2.5 V/0.5 ms, sufficient PNS margin).Results: Quadripolar leads were successfully implanted in 96% of patients. A total of 278 of 299 (93.0%) patients with complete data met the criteria for viable LVPC. With the use of traditional LVPCs, a viable LVPC would have been available (268 of 299 patients; P = 0.002) in significantly fewer patients (89.6%). In any LVPC, PNS was inducible in 65.0% of patients and 22.6% of patients reported PNS during ambulatory 3‐month follow‐up. LVPC reprogramming was performed in 49.8% of patients. PNS inducibility decreased from distal to proximal electrodes, whereas LV pacing thresholds increased from distal to proximal. At prehospital discharge, 5.9 ± 2.8 viable LVPCs were observed, stable during follow‐up. The quadripolar electrode offered significantly more LVPC for LV optimization and PNS avoidance.Conclusion: Quadripolar LV leads yield high numbers of patients with viable LVPCs and alternatives for noninvasive repositioning of LV pacing.Keywords
Funding Information
- St. Jude Medical
This publication has 19 references indexed in Scilit:
- Influence of Pacing Site Characteristics on Response to Cardiac Resynchronization TherapyCirculation: Cardiovascular Imaging, 2013
- 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapyEuropean Heart Journal, 2013
- ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure 2012: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association (HFA) of the ESCEuropean Heart Journal, 2012
- Influence of pacing configurations, body mass index, and position of coronary sinus lead on frequency of phrenic nerve stimulation and pacing thresholds under cardiac resynchronization therapyEP Europace, 2010
- Magnetic resonance imaging and response to cardiac resynchronization therapy: relative merits of left ventricular dyssynchrony and scar tissueEuropean Heart Journal, 2009
- Fluctuation of left ventricular thresholds and required safety margin for left ventricular pacing with cardiac resynchronization therapyEP Europace, 2009
- Impact of viability and scar tissue on response to cardiac resynchronization therapy in ischaemic heart failure patientsEuropean Heart Journal, 2006
- Effect of Posterolateral Scar Tissue on Clinical and Echocardiographic Improvement After Cardiac Resynchronization TherapyCirculation, 2006
- Programmable Multiple Pacing Configurations Help to Overcome High Left Ventricular Pacing Thresholds and Avoid Phrenic Nerve StimulationPacing and Clinical Electrophysiology, 2005
- Long-term retention of cardiac resynchronization therapyJACC, 2004