BACKGROUND AND PURPOSE: Flow-diverter stents are emerging for the endovascular treatment of difficult-to-treat or otherwise untreatable cerebral aneurysms (wide-neck, fusiform, dissecting, blisterlike, or giant). We assessed the clinical safety and efficacy of the Flow-Redirection Endoluminal Device. MATERIALS AND METHODS: This was an institutional review board–approved single-center observational clinical study in 29 patients with 34 aneurysms elected to be treated by endovascular intervention. After providing informed consent, patients were included according to the following criteria: aneurysm fundus-to-neck ratio 4 mm, fusiform, dissecting, or giant aneurysms. The primary end point for clinical safety was the absence of death, absence of major or minor stroke, and absence of transient ischemic attack. The primary end point for treatment efficacy was complete angiographic occlusion according to the O'Kelly Marotta grading scale immediately after the procedure and at follow-up after 3 and 6 months (O'Kelly Marotta D: complete occlusion). RESULTS: The Flow-Redirection Intraluminal Device deployment was technically successful in all cases. In 26/29 (89%) of patients, the primary end point of safety was reached; in the 3 remaining patients, 1 disabling ischemic stroke and 2 minor strokes with complete recovery at follow-up were observed. Angiographic (DSA and MRA) and clinical follow-up were available after 3 months in 29/29 (100%) and after 6 months in 25/29 (86%) patients (after 6 months, only MRA follow-up was performed according to our study protocol and institutional standard). At 3-month follow-up, complete occlusion was reached in 19/34 aneurysms (O'Kelly Marotta D: 19/34; 56%). At 6-month follow-up, aneurysm occlusion was complete in 22/30 aneurysms (O'Kelly Marotta D: 22/30; 73%). CONCLUSIONS: Deployment of the Flow-Redirection Intraluminal Device flow-diverter stent is safe and effective in the treatment of difficult-to-treat or otherwise untreatable intracranial aneurysms.Keywords
This publication has 27 references indexed in Scilit:
- Stent-Supported Aneurysm Coiling: A Literature Survey of Treatment and Follow-UpAmerican Journal of Neuroradiology, 2011
- Hydrogel-coated coils versus bare platinum coils for the endovascular treatment of intracranial aneurysms (HELPS): a randomised controlled trialThe Lancet, 2011
- Better Outcomes with Treatment by Coiling Relative to Clipping of Unruptured Intracranial Aneurysms in the United States, 2001–2008American Journal of Neuroradiology, 2011
- The Pipeline Embolization Device for the Intracranial Treatment of Aneurysms TrialAmerican Journal of Neuroradiology, 2010
- Flow-Diverter Stent for the Endovascular Treatment of Intracranial AneurysmsStroke, 2010
- Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective StudyPLOS ONE, 2010
- Hospital Mortality and Complications of Electively Clipped or Coiled Unruptured Intracranial AneurysmStroke, 2010
- CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trialThe Lancet, 2002
- The “Remodelling Technique” in the Treatment of Wide Neck Intracranial AneurysmsInterventional Neuroradiology, 1997