Simulated optical performance of soft contact lenses on the eye
Open Access
- 14 May 2019
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 14 (5) , e0216484
Abstract
To evaluate the impact of soft contact lens eye-fit on optical power by computational modelling and to produce correction maps for reversing this impact during the design process. Finite element models of spherical and toric hydrogel contact lenses at varying nominal powers of -20 D to +20 D, base curves radii (R1b) of 8.2, 8.5, 8.8 mm, and overall diameters (d3) of 14.5, 15.0, 15.5 mm were generated. Lenses were fitted to computational eye models generated with human eyes’ topography data. Combined eye-lens simulations were run under the boundary conditions of the tears’ surface tension between the contact lens and the eye in addition to the eyelid blink pressure. Lens optical zone power changes were calculated through computational light-ray tracing methods following each simulation. Effective power changes (EPC) were affected negatively for all toric simulated lenses with power varying from -20 D to +20 D. Spherical lenses demonstrated similar behaviour, however with some positive EPC over the power range from -20 D to -10 D for spherical power (SPH) lenses. EPC assessment was between +0.25 D and -0.5 D for most lenses, however, lenses with prescriptions from +10 D to +20 D incurred EPC outside this range. The spherical lenses showed a maximum effective power change of +1.046 ± 0.338 D (Average Eye), and a minimum of -3.278 ± 0.731 D (Steep Eye). Similarly, the toric lenses showed a maximum of +1.501 ± 0.338 D (Average Eye), and a minimum of -3.514 ± 0.731 D (Steep Eye). EPC trends, along with minimum and maximum power, generally increased negatively as nominal lens prescription increased positively. Contact lens base curve selection affected the assessed effective power change for both spherical and toric lenses. The effect from lens total diameter for spherical lenses was less substantial than that for toric lenses. This study considered the impact of soft contact lens design parameters on effective optical power changes (EPC) after eye-fit. Spherical lenses experienced more EPC of clinical significance (>0.25 D) than toric lenses. Both types of lenses, spherical and toric (simple astigmatism), demonstrated similar trends in EPC on fitting from -20 D to +20 D, with lenses in the extremely positive and the extremely negative prescriptions demonstrating the highest EPCs. The lens base curve impacted the extent of EPC observed, with flatter base curves experiencing less power change. Diameter proved to impact toric lenses more than spherical ones, however generally the diameter has less effect on power change than base curve selection.Keywords
Funding Information
- Innovate UK (009521/UVP016)
This publication has 28 references indexed in Scilit:
- Presbyopia and Ocular Conditions Causing Near Vision Impairment in Older Adults From the Brazilian Amazon RegionAmerican Journal of Ophthalmology, 2018
- Theoretical fitting characteristics of typical soft contact lens designsContact Lens and Anterior Eye, 2017
- Comprehensive anterior segment normal values generated by rotating Scheimpflug tomographyJournal of Cataract & Refractive Surgery, 2013
- Contact lens optics and lens designEye, 2007
- Contact lens fitting in high degree myopiaContact Lens and Anterior Eye, 1999
- Surface activity of tear fluid in normal subjectsActa Ophthalmologica Scandinavica, 1998
- On-Eye Power Characteristics of Soft Contact LensesOptometry and Vision Science, 1998
- Shape preserving piecewise cubic interpolationApplied Mathematics-A Journal of Chinese Universities, 1996
- Power and Radius Changes Induced in Soft Contact Lens Systems by FlexureOptometry and Vision Science, 1984
- CORNEAL CURVATURE AND SUPPLEMENTAL POWER EFFECT OF THE BAUSCH AND LOMB SOFLENS® CONTACT LENSOptometry and Vision Science, 1975