Evaluation of two methods for continuous cardiac output assessment during exercise in chronic heart failure patients
Authors:
- Hareld M. C. KempsHKHareld M. C. Kemps
- Eric J. M. ThijssenETEric J. M. Thijssen
- Goof SchepGSGoof Schep
- Boudewijn T. H. M. SleutjesBSBoudewijn T. H. M. Sleutjes
- Wouter R. De VriesWVWouter R. De Vries
- Adwin R. HoogeveenAHAdwin R. Hoogeveen
- Pieter F. F. WijnPWPieter F. F. Wijn
- Pieter A. F. M. DoevendansPDPieter A. F. M. Doevendans
Open Access
- 1 December 2008
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 105 (6) , 1822-1829
Abstract
The purpose of this study was to evaluate the accuracy of two techniques for the continuous assessment of cardiac output in patients with chronic heart failure (CHF): a radial artery pulse contour analysis method that uses an indicator dilution method for calibration (LiDCO) and an impedance cardiography technique (Physioflow), using the Fick method as a reference. Ten male CHF patients (New York Heart Association class II–III) were included. At rest, cardiac output values obtained by LiDCO and Physioflow were compared with those of the direct Fick method. During exercise, the continuous Fick method was used as a reference. Exercise, performed on a cycle ergometer in upright position, consisted of two constant-load tests at 30% and 80% of the ventilatory threshold and a symptom-limited maximal test. Both at rest and during exercise LiDCO showed good agreement with reference values [bias ± limits of agreement (LOA), −1% ± 28% and 2% ± 28%, respectively]. In contrast, Physioflow overestimated reference values both at rest and during exercise (bias ± LOA, 48% ± 60% and 48% ± 52%, respectively). Exercise-related within-patient changes of cardiac output, expressed as a percent change, showed for both techniques clinically acceptable agreement with reference values (bias ± LOA: 2% ± 26% for LiDCO, and −2% ± 36% for Physioflow, respectively). In conclusion, although the limits of agreement with the Fick method are pretty broad, LiDCO provides accurate measurements of cardiac output during rest and exercise in CHF patients. Although Physioflow overestimates cardiac output, this method may still be useful to estimate relative changes during exercise.Keywords
This publication has 41 references indexed in Scilit:
- Non-invasive measurement of cardiac output in patients with chronic heart failureBlood Pressure Monitoring, 2004
- Reliability and Validity of Measures of Cardiac Output During Incremental to Maximal Aerobic ExerciseSports Medicine, 1999
- Reliability and Validity of Measures of Cardiac Output During Incremental to Maximal Aerobic ExerciseSports Medicine, 1999
- Clinical Application of Cardiac Output During Ramp Exercise Calculated Using the Fick EquationJapanese Circulation Journal, 1997
- Comparison of impedance cardiography with thermodilution and direct fick methods for noninvasive measurement of stroke volume and cardiac output during incremental exercise in patients with ischemic cardiomyopathyThe American Journal of Cardiology, 1996
- Beat-to-beat evaluation of cardiac function during recovery from upright bicycle exercise in patients with coronary artery diseaseAmerican Heart Journal, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Measurement of cardiac output during exercise by the thermodilution and direct fick techniques in patients with chronic congestive heart failureThe American Journal of Cardiology, 1985
- Characteristics of vascular hydraulic load in patients with heart failure.Circulation, 1985
- Accuracy of cardiac output, oxygen uptake and arteriovenous oxygen difference at rest, during exercise and after vasodilator therapy in patients with severe, chronic heart failureThe American Journal of Cardiology, 1982