Non-Invasive Wearable Patch Utilizing Seismocardiography for Peri-Operative Use in Surgical Patients
- 11 May 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Biomedical and Health Informatics
- Vol. 25 (5) , 1572-1582
Abstract
Objective: Optimizing peri-operative fluid management has been shown to improve patient outcomes and the use of stroke volume (SV) measurement has become an accepted tool to guide fluid therapy. The Transesophageal Doppler (TED) is a validated, minimally invasive device that allows clinical assessment of SV. Unfortunately, the use of the TED is restricted to the intra-operative setting in anesthetized patients and requires constant supervision and periodic adjustment for accurate signal quality. However, post-operative fluid management is also vital for improved outcomes. Currently, there is no device regularly used in clinics that can track patient's SV continuously and non-invasively both during and after surgery. Methods: In this paper, we propose the use of a wearable patch mounted on the mid-sternum, which captures the seismocardiogram (SCG) and electrocardiogram (ECG) signals continuously to predict SV in patients undergoing major surgery. In a study of 12 patients, hemodynamic data was recorded simultaneously using the TED and wearable patch. Signal processing and regression techniques were used to derive SV from the signals (SCG and ECG) captured by the wearable patch and compare it to values obtained by the TED. Results: The results showed that the combination of SCG and ECG contains substantial information regarding SV, resulting in a correlation and median absolute error between the predicted and reference SV values of 0.81 and 7.56 mL, respectively. Significance: This work shows promise for the proposed wearable-based methodology to be used as an alternative to TED for continuous patient monitoring and guiding peri-operative fluid management.Keywords
Funding Information
- National Institutes of Health (1R01HL130619-A1)
This publication has 44 references indexed in Scilit:
- Extracting respiratory information from seismocardiogram signals acquired on the chest using a miniature accelerometerPhysiological Measurement, 2012
- Impedance cardiography versus invasive measurements of stroke volume index in patients with chronic heart failureInternational Journal of Cardiology, 2010
- Systematic review of the literature for the use of oesophageal Doppler monitor for fluid replacement in major abdominal surgeryAnaesthesia, 2007
- Doppler-guided intra-operative fluid management during major abdominal surgery: systematic review and meta-analysisInternational Journal Of Clinical Practice, 2007
- Modelflow: a new method for noninvasive assessment of cardiac output in pregnant womenAmerican Journal of Obstetrics and Gynecology, 2007
- Randomized clinical trial assessing the effect of Doppler-optimized fluid management on outcome after elective colorectal resectionBritish Journal of Surgery, 2006
- Newer Classification and Regression Tree Techniques: Bagging and Random Forests for Ecological PredictionEcosystems, 2006
- Cubic-spline interpolation. 1IEEE Instrumentation & Measurement Magazine, 2001
- Interaction between respiratory and RR interval oscillations at low frequenciesJournal of the Autonomic Nervous System, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986